FDA Adverse Event Other Summary report: N

MAQUET SA

MDR report key: 863699 · Received September 26, 2006

Report

Report Number
9710055-2006-00005
Event Type
Other
Date Received
September 26, 2006
Report Date
August 21, 2006
Manufacturer
MAQUET S.A.
Product Code
FTD
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
ND, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

MANUFACTURER DETERMINED THAT INCIDENT OCCURRED DUE TO LACK OF MAINTENANCE ON THE UNIT. MAINTENANCE INFORMATION FOR THE UNIT PROVIDED IN THE USER'S MANUAL ADVISES THAT THE CONDITION OF ALL PAINTED PARTS, ESPECIALLY THOSE WHO ARE SUBECT TO IMPACT FROM OTHER EQUIPMENT SHOULD BE CHECKED PERIODICALLY. INFORMATION ON THE AVAILABILITY OF TOUCH UP PAINT WAS PROVDED. GETTING USA, INC SUBMITS THIS REPORT ON BEHALF OF THE DEVICE MANFUACTURING FACILITY. MAQUET PROVIDES PRODUCT FAILURE INVESTIGATION, ANALYSIS AND RESOLUTION FOR THE DEVICE DESCIRBED IN THIS REPORT.

Description of Event or Problem · 1

PAINT CHIPS FROM SURGICAL LIGHT FELL INTO OPEN ORTHOPEDIC WOUND DURING SURGERY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MAQUET SA SURGICAL LAMP FTD MAQUET S.A. PRC *

Patients

Seq Age Sex Outcome Treatment
1 YR