640G INSULIN PUMP MMT-1711K
Report
- Report Number
- 2032227-2019-05645
- Event Type
- Malfunction
- Date Received
- May 23, 2019
- Date of Event
- November 8, 2018
- Report Date
- May 17, 2022
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- UDI-DI
- 00643169999763
- Removal / Correction Number
- Z-0958-2020
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NL
- Reporter Occupation
- OTHER
- Health Professional
- N
Narratives
THIS REPORT IS PART OF A RETROSPECTIVE REVIEW AND REMEDIATION EFFORTS IN RESPONSE TO A WARNING LETTER. UPDATED H9: Z-0958-2020. MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT.
UNIT RECEIVED WITH MISSING RETAINER RING AND RESERVOIR TUBE LIP O-RING. UNABLE TO PERFORM DISPLACEMENT TEST AND P-CAP / RESERVOIR WILL NOT LOCK PROPERLY DUE TO MISSING RETAINER. UNIT RECEIVED WITH CRACKED CASE ON THE BACK AT THE BATTERY TUBE AREA AND BATTERY TUBE THREADS. THE INSULIN PUMP INVOLVED IN THIS EVENT IS THE 640G INSULIN INFUSION PUMP, WHICH IS NOT MARKETED IN THE UNITED STATES. HOWEVER, THE DEVICE IS SIMILAR TO THE PARADIGM REAL-TIME INSULIN INFUSION PUMP, WHICH IS MARKETED IN THE UNITED STATES.
ON (B)(6) 2018, 16:13:02 PST ICE BATCH USER (BATCH_ICE), (B)(6), COMPLAINT #: (B)(4), COMPLAINT STATUS: IN PROCESS, MDT INITIAL CONTACT: (B)(6), TAKEN BY: (B)(6). PUMP REPORTS A CRACK NEAR BATTERY COMPARTMENT. EXCHANGE VIA EOC. COUNTRY: (B)(6), INPUT DATE: (B)(6) 2018, WARRANTY START: 03/01/2016, WARRANTY END: 02/29/2020. BATCH - (B)(4), (B)(6) 2018 17:10:08, PST ICE BATCH USER (BATCH_ICE), (B)(6), COMPLAINT: (B)(4), COMPLAINT STATUS: IN PROCESS, MDT INITIAL CONTACT: (B)(6), TAKEN BY: (B)(6). PUMP REPORTS A CRACK NEAR BATTERY COMPARTMENT. EXCHANGE VIA EOC. COUNTRY: (B)(6), INPUT DATE: (B)(6) 2018, WARRANTY START: 03/01/2016, WARRANTY END: 02/29/2020. BATCH - (B)(4), (B)(6) 2018 16:38:40, PST ICE BATCH USER (BATCH_ICE), (B)(6), COMPLAINT: (B)(4), COMPLAINT STATUS: IN PROCESS, MDT INITIAL CONTACT: (B)(6), TAKEN BY: (B)(6). PUMP REPORTS A CRACK NEAR BATTERY COMPARTMENT. EXCHANGE VIA EOC. MASTER BATCH NOT FILLED IN EQUIPM NR SAP. PLEASE CREATE MANUALLY AN RMA. COUNTRY: (B)(6), INPUT DATE: (B)(6) 2018, WARRANTY START: 03/01/2016, WARRANTY END: 02/29/2020. BATCH - (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 431861 | 640G INSULIN PUMP MMT-1711K | PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-1711K | 00643169999763 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |