LIGACLIP ENDOSCOPIC MULTIPLE CLIP APPLIER
Report
- Report Number
- 1527736-1997-00703
- Event Type
- Malfunction
- Date Received
- April 21, 1997
- Date of Event
- March 24, 1997
- Report Date
- April 21, 1997
- Manufacturer
- ETHICON ENDO-SURGERY, INC.
- Product Code
- GDO
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- PHYSICIAN
Narratives
FACILITY EXPERIENCED AN EVENT WITH YOUR LIGACLIP ENDOSCOPIC MULTIPLE CLIP APPLIER WHILE PERFORMING A LAP CHOLECYSTECTOMY. THE PRODUCT COMPLAINT ANALYSIS TEAM HAS COMPLETED ITS INVESTIGATION OF THE DEVICE WHICH WAS RETURNED TO CO WITH PRODUCT INQUIRY #971858. THE RESULTS OF THE INVESTIGATION CONDUCTED BY THE APPROPRIATE ENGINEERS AND TECHNICIANS ARE LISTED BELOW. VISUAL INSPECTIONS & RESULTS: CLIP IN JAWS NO, DAMAGED JAWS NO, DAMAGED CUTTER N/A, DAMAGED FEED BAR NO, DAMAGED FLOOR NO, DAMAGED HANDLE SHROUDS NO, DAMAGED OTHER NO, DAMAGED TIP SHROUDS NO, DAMGED TRIGGER NO, DAMAGED TUBE NO, DAMAGED WELD SEAMS NO. FUNCTIONAL TESTS & RESULTS: ANTIBACKUP FUNCTIONAL YES, FIRING: FEED CONFORM YES, FIRING: FORM CONFORM YES, JAWS: HOLD CLIP YES, JAWS: INSIDE WIDTH AT TIPS .195, LOCKOUT FUNCTIONAL YES, NUMBER OF CLIPS FED AND FORMED 10. ANALYSIS CONCLUSION: BASED ON THE INFO RECEIVED AND THE VISUAL AND FUNCTIONAL RESULTS, NO CONCLUSION COULD BE REACHED AS TO WHAT MAY HAVE CAUSED THE REPORTED INCIDENT. THE INSTGRUMENT FED, FIRED AND FORMED THE REMAINING CLIPS AS DESIGNED. THE REPORTED INCIDENT COULD NOT BE RECREATED. EACH INSTRUMENT IS EVALUATED DURING THE ASSEMBLY PROCESS TO ENSURE THE CLIPS FEED AND FORM PROPERLY. CO STRIVES TO UNDERSTAND EACH INCIDENT AS IT OCCURS IN ORDER TO CONTINUOUSLY IMPROVE THE PRODUCTS.
THE DEVICE WAS USED DURING A GALL BLADDER PROCEDURE. IT WAS REPORTED BY THE AFFILIATE THAT THE CLIPS ARE MALFORMED. THERE WAS NO PATIENT CONSEQUENCE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LIGACLIP ENDOSCOPIC MULTIPLE CLIP APPLIER | ENDOSCOPIC CLIP APPLIER | GDO | ETHICON ENDO-SURGERY, INC. | NA | J4239A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNKNOWN | Other |