FDA Adverse Event Malfunction Summary report: N

BD INTIMA-II¿ CLOSED IV CATHETER SYSTEM

MDR report key: 8627887 · Received May 21, 2019

Report

Report Number
3006948883-2019-00387
Event Type
Malfunction
Date Received
May 21, 2019
Date of Event
April 29, 2019
Report Date
June 18, 2019
Manufacturer
BD (SUZHOU)
Product Code
FOZ
PMA / PMN Number
N/A
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION SUMMARY: A DEVICE HISTORY REVIEW WAS CONDUCTED FOR LOT NUMBER 7234053. OUR RECORDS SHOW THAT THIS IS THE ONLY INSTANCE OF THIS ISSUE OCCURRING IN THIS PRODUCTION BATCH. ACCORDING TO THE SAMPLING PLAN APPLIED FOR PRODUCT PERFORMANCE, THIS LOT WAS ACCEPTED AND RELEASED WITHOUT DEFECTS BEING NOTED DURING THE FINAL ASSEMBLY OR VISUAL INSPECTIONS. UNFORTUNATELY A SAMPLE COULD NOT BE OBTAINED FOR EVALUATION AND TESTING. WITHOUT THE ABILITY TO INVESTIGATE THE AFFECTED UNIT OUR QUALITY ENGINEERS WERE UNABLE TO DETERMINE THE ROOT CAUSE FOR THIS COMPLAINT.

Description of Event or Problem · 0

IT WAS REPORTED THAT CLAMP NOT TIGHT AND THERE WAS LEAKAGE WITH A BD INTIMA-II¿ CLOSED IV CATHETER SYSTEM. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM CHINESE TO ENGLISH: NURSE FEEDBACK CLAMP WASN'T CLAMPED TIGHTLY, EASY TO OPEN.

Additional Manufacturer Narrative · 1

INITIAL REPORTER PHONE #: UNKNOWN. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED THAT CLAMP NOT TIGHT AND THERE WAS LEAKAGE WITH A BD INTIMA-II¿ CLOSED IV CATHETER SYSTEM. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM CHINESE TO ENGLISH: NURSE FEEDBACK CLAMP WASN'T CLAMPED TIGHTLY, EASY TO OPEN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
422967 BD INTIMA-II¿ CLOSED IV CATHETER SYSTEM INTRAVASCULAR CATHETER FOZ BD (SUZHOU) 7234053

Patients

Seq Age Sex Outcome Treatment
1 Other