EVIS EXERAIII BRONCHOVIDEOSCOPE
Report
- Report Number
- 8010047-2019-01984
- Event Type
- Malfunction
- Date Received
- May 21, 2019
- Date of Event
- April 12, 2019
- Report Date
- July 10, 2019
- Manufacturer
- OLYMPUS MEDICAL SYSTEMS CORP.
- Product Code
- EOQ
- PMA / PMN Number
- K121959
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AS
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
THIS SUPPLEMENTAL REPORT IS BEING SUBMITTED TO PROVIDE THE DEVICE EVALUATION RESULT. THE SUBJECT DEVICE HAS NOT BEEN RETURNED TO OLYMPUS MEDICAL SYSTEMS CORP. (OMSC) BUT WAS RETURNED TO OLYMPUS AUSTRALIA.(OAZ) FOR EVALUATION. IN THE EVALUATION, OAZ CONFIRMED A PART OF THE LIGHT GUIDE OF THE SUBJECT DEVICE WAS LOOSE. THE EXACT CAUSE OF THE REPORTED EVENT COULD NOT BE CONCLUSIVELY DETERMINED AT THIS TIME. IF ADDITIONAL INFORMATION IS RECEIVED, THIS REPORT WILL BE SUPPLEMENTED.
THE SUBJECT DEVICE HAS NOT BEEN RETURNED TO OMSC BUT WAS RETURNED TO OLYMPUS (B)(4) FOR EVALUATION. IN THE EVALUATION, (B)(4) CONFIRMED THE LEAKAGE FROM THE INSTRUMENT CHANNEL OF THE SUBJECT DEVICE. OMSC REVIEWED THE MANUFACTURING HISTORY (DHR) OF THE SUBJECT DEVICE AND CONFIRMED NO IRREGULARITY. THE EXACT CAUSE OF THE REPORTED EVENT COULD NOT BE CONCLUSIVELY DETERMINED AT THIS TIME. IF ADDITIONAL INFORMATION IS RECEIVED, THIS REPORT WILL BE SUPPLEMENTED.
OLYMPUS MEDICAL SYSTEMS CORP. (OMSC) WAS INFORMED THAT AS A RESULT OF ROUTINE MICROBIOLOGICAL TESTING BY THE USER FACILITY, THE SUBJECT DEVICE TESTED POSITIVE FOR UNSPECIFIED MICROBES. THE USER FACILITY DID NOT PROVIDE OTHER DETAILED INFORMATION SUCH AS THE NUMBER AND THE TYPE OF MICROBES. THE DEVICE HAD BEEN MANUALLY REPROCESSED. THERE WAS NO REPORT OF INFECTION ASSOCIATED WITH THIS REPORT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 423669 | EVIS EXERAIII BRONCHOVIDEOSCOPE | BRONCHOVIDEOSCOPE | EOQ | OLYMPUS MEDICAL SYSTEMS CORP. | BF-1TH190 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |