FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-723NAP

MDR report key: 8623280 · Received May 20, 2019

Report

Report Number
2032227-2019-04372
Event Type
Malfunction
Date Received
May 20, 2019
Date of Event
May 6, 2019
Report Date
June 17, 2019
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
UDI-DI
00643169507241
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Additional Manufacturer Narrative · 1

DEVICE RECEIVED WITH INTERMITTENT BUTTON RESPONSE DUE TO FLATTENED ESC AND ACT BUTTON DOME SWITCH. NO BUTTON ERROR ALARM DURING TESTING. NO UNLOCKED J2 OR LCD KEYPAD CONNECTOR. DEVICE PASSED SELF TEST.

Description of Event or Problem · 1

THE CUSTOMER REPORTED VIA PHONE CALL THAT THE INSULIN PUMP¿S BUTTON HAD HARD TO PRESS. CUSTOMER¿S BLOOD GLUCOSE LEVEL WAS 106 MG/DL. CUSTOMER REPORTED THAT ACT BUTTON WAS HARD TO PRESS. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.