FDA Adverse Event Injury Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-723LNAS

MDR report key: 8622310 · Received May 18, 2019

Report

Report Number
3004209178-2019-99044
Event Type
Injury
Date Received
May 18, 2019
Date of Event
May 15, 2019
Report Date
July 1, 2019
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
UDI-DI
00643169513914
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AR, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DEVICE PASSED THE REWIND, BASIC OCCLUSION, PRIME AND DISPLACEMENT TEST. DEVICE RECEIVED WITH STUCK MOTOR ERROR ALARM LOOP DURING BOLUS DELIVERY. THE MOTOR WAS TESTED OUTSIDE OF THE DEVICE AND PASSED. THE MOTOR MAY HAVE HAD AND INTERMITTENT FAILURE THAT WAS NOT DETECTED DURING OUR TESTING.

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Description of Event or Problem · 1

CUSTOMER¿S PARENTS REPORTED VIA PHONE CALL THAT HER DAUGHTER EXPERIENCED HIGH BLOOD GLUCOSE. CUSTOMER¿S CURRENT BLOOD GLUCOSE WAS 463 MG/DL AT THE TIME OF INCIDENT. CUSTOMER ALSO STATED THAT THE INSULIN PUMP HAD MOTOR ERROR. CUSTOMER WAS ABLE TO SUCCESSFULLY CLEAR THE ALARM AND ABLE TO COMPLETE INSULIN PUMP REWIND. DRIVE SUPPORT CAP WAS NORMAL AND CUSTOMER DID NOT REPORT MOTOR POSITION ENCODER ALARM. THE CUSTOMER WAS ADVISED THAT THE INSULIN PUMP NEEDED TO BE REPLACED AND WAS ADVISED TO DISCONTINUE USE OF THE INSULIN PUMP AND TO REVERT TO THE BACKUP PLAN. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
415557 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-723LNAS PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-723LNAS A5723LNASJ 00643169513914

Patients

Seq Age Sex Outcome Treatment
1 17 YR Other