SPECTRUM INFUSION PUMP
Report
- Report Number
- 1314492-2019-01730
- Event Type
- Malfunction
- Date Received
- May 17, 2019
- Date of Event
- April 25, 2019
- Report Date
- May 17, 2019
- Manufacturer
- BAXTER HEALTHCARE - MEDINA
- Product Code
- FRN
- UDI-DI
- 00085412498683
- PMA / PMN Number
- K133801
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
BAXTER RECEIVED AND EVALUATED THE DEVICE. VISUAL INSPECTION DID NOT IDENTIFY ANY ABNORMALITIES THAT COULD HAVE CONTRIBUTED TO THE REPORTED CONDITION. A DEVICE HISTORY REVIEW REVEALED NO ISSUES THAT COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED ISSUE. THE REPORTED CONDITION WAS VERIFIED. THE DEVICE WAS FOUND OUT OF SPECIFICATION IN RELATION TO THE REPORTED DOESN'T DETECT OPEN DOOR WHICH WAS REPRODUCED DURING EVALUATION AND FOUND TO BE CAUSED BY A FAILED UPPER LATCH SWITCH. THE UPPER AUXILIARY WAS REPLACED TO CORRECT THIS CONDITION. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
IT WAS REPORTED THAT A SPECTRUM PUMP DID NOT DETECT OPEN DOOR DURING PROGRAMMING OR SETUP. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 413601 | SPECTRUM INFUSION PUMP | INFUSION PUMP | FRN | BAXTER HEALTHCARE - MEDINA | NA | NA | 00085412498683 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |