FDA Adverse Event Malfunction Summary report: N

R-PILOT, 6X, STERILE (25MM)

MDR report key: 8621690 · Received May 17, 2019

Report

Report Number
9611053-2019-00137
Event Type
Malfunction
Date Received
May 17, 2019
Report Date
May 17, 2019
Manufacturer
DENTSPLY VDW GMBH
Product Code
EKS
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
DENTIST

Narratives

Additional Manufacturer Narrative · 1

THERE HAS BEEN A PREVIOUS REPORT RECEIVED WITH A SIMILAR DEVICE WHERE THIS MALFUNCTION RESULTED IN A SERIOUS INJURY. THEREFORE, IT MUST BE PRESUMED THAT RECURRENCE OF THIS MALFUNCTION COULD POSSIBLY CAUSE OR CONTRIBUTE TO A SERIOUS INJURY OR REQUIRE MEDICAL OR SURGICAL INTERVENTION TO PRECLUDE SUCH. AS SUCH, THIS EVENT IS REPORTABLE PER 21CFR PART 803. THE INVOLVED R-PILOT 25MM IS ACTUALLY BROKEN AT THE TIP OF THE ACTIVE PART (TORQUE). NO MATERIAL DEFECT WAS FOUND DURING ANALYSIS OF THE RUPTURE PATTERN. UNUSED PRODUCT HAS BEEN EVALUATED ACCORDING TO OUR PRESCRIPTIONS AND WAS FOUND IN COMPLIANCE WITH SPECIFICATIONS (MEASURES, TORQUE TEST). NOTHING UNUSUAL TO REPORT WAS FOUND DURING DHRS REVIEW (BATCHES #1555813, 1555815). ROOT CAUSES ARE NOT IDENTIFIED. THIS KIND OF EVENT IS TRACKED AND WE MONITOR THE TREND.

Description of Event or Problem · 1

IN THIS EVENT IT WAS REPORTED THAT A R-PILOT BROKE DURING USE. THE BROKEN PART REMAINED IN ROOT CANAL AND TOOTH WAS FILLED WITH BROKEN PART.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
414976 R-PILOT, 6X, STERILE (25MM) FILE, PULP CANAL, ENDODONTIC EKS DENTSPLY VDW GMBH NA 277972

Patients

Seq Age Sex Outcome Treatment
1