FDA Adverse Event Malfunction Summary report: N

SOLIA S 53

MDR report key: 8621555 · Received May 17, 2019

Report

Report Number
1028232-2019-02031
Event Type
Malfunction
Date Received
May 17, 2019
Date of Event
May 7, 2019
Report Date
May 10, 2019
Manufacturer
BIOTRONIK SE & CO. KG
Product Code
NVN
UDI-DI
04035479118266
PMA / PMN Number
P950037
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

UPON RECEIPT, THE LEAD UNDER COMPLAINT WAS SUBJECTED TO AN EXTENSIVE ANALYSIS. THE PERFORMANCE OF THE DEVICE WAS SCRUTINIZED, INCLUDING A VISUAL AND ELECTRICAL INSPECTION. THE ANALYSIS OF THE LEAD DEMONSTRATED A DAMAGED INSULATION APPROXIMATELY 24.5CM AND 25.5CM DISTAL TO THE IS-1 CONNECTOR PIN. AT THESE POINTS, THE LEAD BODY WAS FOUND SQUEEZED AND DEFORMED. BASED ON THE CHARACTERISTICS AS WELL AS THE LOCATION OF THE DAMAGE, IT IS REASONABLE TO ASSUME THAT THE LEAD HAD BEEN SUBJECT TO EXCESSIVE MECHANICAL FORCES AS THE RESULT OF CLAVICULAR FIRST RIB ENTRAPMENT. THE MANUFACTURING PROCESS FOR THIS LEAD WAS RE-INVESTIGATED. ALL PRODUCTION STEPS HAD BEEN PERFORMED ACCORDINGLY. THERE WAS NO SIGN OF ANY INCONSISTENCY DURING THE MANUFACTURING PROCESS. IN CONCLUSION, THE ANALYSIS DID NOT REVEAL ANY SIGN OF A MATERIAL OR MANUFACTURING PROBLEM.

Description of Event or Problem · 1

THIS SYSTEM WAS REMOVED BECAUSE THE RA AND RV LEADS FLIPPED TO UNIPOLAR AND THERE WAS LOSS OF CAPTURE IN UNIPOLAR. THE PHYSICIAN CHOSE TO REPLACE THE WHOLE SYSTEM.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
413858 SOLIA S 53 PACING LEAD NVN BIOTRONIK SE & CO. KG 377177 04035479118266

Patients

Seq Age Sex Outcome Treatment
1 58 YR