FDA Adverse Event Malfunction Summary report: N

DIMENSION VISTA®

MDR report key: 8621377 · Received May 17, 2019

Report

Report Number
2517506-2019-00203
Event Type
Malfunction
Date Received
May 17, 2019
Date of Event
April 23, 2019
Report Date
May 17, 2019
Manufacturer
SIEMENS HEALTHCARE DIAGNOSTICS INC
Product Code
JJE
PMA / PMN Number
K143720
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SC, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE CUSTOMER CONTACTED THE SIEMENS HEALTHCARE DIAGNOSTICS CUSTOMER CARE CENTER (CCC) AND REPORTED THAT A DISCORDANT, FALSELY DEPRESSED VANCOMYCIN (VANC) RESULT WAS OBTAINED ON A DIMENSION VISTA 1500 SYSTEM. SIEMENS HEALTHCARE DIAGNOSTICS HEADQUARTERS SUPPORT CENTER (HSC) COMPLETED THE INVESTIGATION OF THE EVENT. HSC HAS REVIEWED THE INFORMATION PROVIDED AND THE INSTRUMENT DATA. CALIBRATION AND QUALITY CONTROL DATA WERE WITHIN CONFORMANCE. THE CUSTOMER COMPLETED BASIC TROUBLESHOOTING AT THE REQUEST OF THE CUSTOMER CARE CENTER (CCC). THE CUSTOMER DECLINED ADDITIONAL INSTRUMENT HARDWARE SUPPORT. THERE IS NO EVIDENCE OF A PRODUCT NONCONFORMANCE. THE CAUSE OF THE EVENT IS UNKNOWN. THE DEVICE IS PERFORMING WITHIN SPECIFICATIONS. NO FURTHER EVALUATION IS REQUIRED.

Description of Event or Problem · 1

A DISCORDANT FALSELY DEPRESSED VANCOMYCIN (VANC) RESULT WAS OBTAINED ON A PATIENT SAMPLE ON A DIMENSION VISTA 1500 INSTRUMENT. THE DISCORDANT RESULT WAS REPORTED TO THE PHYSICIAN WHO QUESTIONED THE RESULT. A NEW SAMPLE WAS PROCESSED THE SAME DAY AND A HIGHER RESULT WAS OBTAINED AND REPORTED. THE ORIGINAL SAMPLE WAS REPROCESSED IN DUPLICATE ON THE ORIGINAL DIMENSION VISTA AT A LATER DATE AND HIGHER RESULTS WERE OBTAINED. A CORRECTED RESULT WAS REPORTED. THERE WERE NO REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT FALSELY DEPRESSED VANC RESULT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
414151 DIMENSION VISTA® DIMENSION VISTA®1500 SYSTEM JJE SIEMENS HEALTHCARE DIAGNOSTICS INC DIMENSION VISTA®1500 SYSTEM

Patients

Seq Age Sex Outcome Treatment
1