FDA Adverse Event Malfunction Summary report: N

EXCEL T-HANDLE

MDR report key: 8621300 · Received May 17, 2019

Report

Report Number
1818910-2019-93930
Event Type
Malfunction
Date Received
May 17, 2019
Date of Event
April 29, 2019
Report Date
April 30, 2019
Manufacturer
DEPUY ORTHOPAEDICS, INC. 1818910
Product Code
LXH
UDI-DI
10603295079941
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AU
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISION OF 21 CFR, PART 803. THE REPORT MAY BE BASED ON THE INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. PRODUCT COMPLAINT # (B)(4). INVESTIGATION SUMMARY: EXAMINATION OF THE RETURNED DEVICE CONFIRMS THE COMPLAINT OF HANDLE DAMAGE, THE HANDLE IS CRACKED. ROOT CAUSE DEVICE WORN FROM NORMAL USE AND SERVICING. DEPUY SYNTHES CONSIDERS THE INVESTIGATION CLOSED AT THIS TIME. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THE INFORMATION WILL BE REVIEWED AND THE INVESTIGATION MAY BE RE-OPENED AS NECESSARY. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 1

PRODUCT COMPLAINT # (B)(4). IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 1

(B)(4). IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Description of Event or Problem · 1

HANDLE OF EXCEL T HANDLE HAS STARTED TO SPLIT AT THE HANDLE/SHAFT INTERSECTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
414440 EXCEL T-HANDLE HIP INSTRUMENTS : HANDLES LXH DEPUY ORTHOPAEDICS, INC. 1818910 SO2036771 10603295079941

Patients

Seq Age Sex Outcome Treatment
1