FORTIFY VR, U1.6 SJ4 US
Report
- Report Number
- 2017865-2019-08250
- Event Type
- Injury
- Date Received
- May 17, 2019
- Date of Event
- April 30, 2019
- Report Date
- September 4, 2019
- Manufacturer
- ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
- Product Code
- NVZ
- UDI-DI
- 05414734504348
- PMA / PMN Number
- P910023
- Removal / Correction Number
- Z-0003-2018;Z-0115-2017
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TH
- Reporter Occupation
- PHYSICIAN
Narratives
THE RESULTS OF THE INVESTIGATION ARE INCONCLUSIVE SINCE THE DEVICE WAS NOT RETURNED FOR ANALYSIS. BASED ON THE INFORMATION RECEIVED, THE CAUSE OF THE REPORTED INCIDENT COULD NOT BE CONCLUSIVELY DETERMINED. THE DEVICE IS INCLUDED IN THE BATTERY PERFORMANCE ALERT ADVISORY ISSUED BY ABBOTT ON 28 AUGUST 2017.
PREMATURE BATTERY DEPLETION WAS CONFIRMED BY ANALYSIS. NO SOURCES OF HIGH CURRENT WERE NOTED. THE CAUSE OF THE PREMATURE BATTERY DEPLETION WAS CONSISTENT WITH LITHIUM (LI) CLUSTER FORMATION. FROM THESE ANALYSES, IN THE ABSENCE OF HIGH CURRENT DRAW, IT IS PROBABLE THAT THE PREMATURE BATTERY DEPLETION WAS CAUSED BY A LITHIUM CLUSTER INDUCED SHORT CIRCUIT. LI CLUSTERS ARE A KNOWN DEPLETION MECHANISM FOR THESE ADVISORY PRODUCTS THAT HAS BEEN INVESTIGATED AND ASSOCIATED WITH A FIELD ACTION IN OCTOBER 2016.
FOLLOWING THE BATTERY PERFORMANCE ALERT (BPA) ADVISORY, A BPA WAS RECEIVED BY THE CLINICIAN AND THE DEVICE WAS EXPLANTED. PREMATURE BATTERY DEPLETION WAS ALSO SUSPECTED. THE PATIENT EXPERIENCED NO ADVERSE CONSEQUENCES AND WAS STABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 415142 | FORTIFY VR, U1.6 SJ4 US | IMPLANTABLE CARDIOVERTER DEFIBRILLATOR | NVZ | ST. JUDE MEDICAL, INC.(CRM-SYLMAR) | CD1231-40Q | 3965352 | 05414734504348 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 61 YR | Required Intervention |