Narratives

Additional Manufacturer Narrative · 0

THIS CORRECTION MDR IS BEING SUBMITTED TO ADDRESS INACCURATE INFORMATION IN SECTION F ON THE XML VERSION OF THE PREVIOUS SUBMISSION CAUSED BY A SOFTWARE GLITCH IN THE TRANSMISSION PROCESS. THE INTERNAL NON-CONFORMANCE NUMBERS FOR THIS SOFTWARE ISSUE ARE (B)(4) AND CAPA-010215.

Additional Manufacturer Narrative · 1

(B)(4). THE PATIENT INTERFACE (PI) SUCTION RING MAY LOSE SUCTION DURING A PROCEDURE. LABEL COPY STATES CORNEAL FIXATION VACUUM LOSS CAN OCCUR. THERE ARE SEVERAL FACTORS THAT MAY CONTRIBUTE TO SUCTION ISSUES SUCH AS DOCTOR¿S TECHNIQUE IN APPLYING THE SUCTION RING TO THE CORNEA, DOCTOR¿S TECHNIQUE IN SQUEEZING THE PI CLIP TO SECURE THE SUCTION RING TO THE PI CONE AND PATIENT ANATOMY AFFECTING THE INTERFACE BETWEEN THE PATIENT¿S CORNEA AND THE SUCTION RING. A MANUFACTURER RECORD REVIEW RELATED TO THE DEVICE INCLUDING DEVICE HISTORY RECORD WILL BE PERFORMED. UPON COMPLETION OF THIS REVIEW, IF THERE IS ANY FURTHER RELEVANT INFORMATION OBTAINED THAT CHANGES THE FACTS AND/OR CONCLUSION, A SUPPLEMENTAL MEDWATCH WILL BE FILED. ALL PERTINENT INFORMATION AVAILABLE TO JOHNSON & JOHNSON SURGICAL VISION, INC. HAS BEEN SUBMITTED.

Description of Event or Problem · 1

THE SURGERY CENTER REPORTED SUCTION LOSS DURING SURGERY. THE PATIENT''S PROCEDURE IS TO BE CONVERTED TO PRK (PHOTOREFRACTIVE KERATECTOMY). THIS REPORT IS FOR PATIENT 2 OF 2.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
414911 INTRALASE PATIENT INTERFACE HNO JOHNSON & JOHNSON SURGICAL VISION, INC. PI-RET 60155494 15050474534688

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention FEMTO LASER| FEMTO LASER