STYLE 68 SALINE FILLED BREAST IMPLANT
Report
- Report Number
- 9617229-2019-03261
- Event Type
- Injury
- Date Received
- May 17, 2019
- Date of Event
- April 17, 2019
- Report Date
- September 4, 2019
- Manufacturer
- ALLERGAN (COSTA RICA)
- Product Code
- FWM
- PMA / PMN Number
- P990074
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WA
- Reporter Occupation
- PHYSICIAN
Narratives
ANALYSIS OF THE RETURNED DEVICE IDENTIFIED: WEAR ABRASION, CREASE FOLD, OPENING ON POSTERIOR BROWN AND YELLOW PARTICLES DEVICE INNER SURFACE AND RED PARTICLES IN FILL CHANNEL. LEAK TEST AND MICROSCOPIC ANALYSIS WAS PERFORMED WHICH IDENTIFIED: OPENING CREASE SHARP, OPENING CREASE SMOOTH AND OBSERVED VOID >1 MM IN NON-TEXTURED, VALVE-TO SHELL-BOND AREA. THE FILL TEST INSPECTION WAS PERFORMED, THE RESULT IS NO BLOCKAGE. BASED ON THE DEVICE ANALYSIS THE FINAL ASSESSMENT IS: AN OPENING CREASE SHARP ON ANTERIOR DUE TO FOLD FLAW OPENING, AND A SMOOTH CREASE OPENING ON POSTERIOR DUE TO FOLD FLAW OPENING.
A REVIEW OF THE DEVICE HISTORY RECORD HAS BEEN COMPLETED. NO DEVIATIONS OR NON-CONFORMANCES NOTED. FURTHER INFORMATION FROM THE REPORTER REGARDING EVENT, PRODUCT, OR PATIENT DETAILS HAS BEEN REQUESTED. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME. THE REASON FOR REOPERATION IS DEFLATION. THIS IS A KNOWN POTENTIAL ADVERSE EVENT ADDRESSED IN THE PRODUCT LABELING.
HEALTHCARE PROFESSIONAL REPORTED A RIGHT SIDE DEFLATION. THE DEVICE WAS EXPLANTED.
HEALTHCARE PROFESSIONAL REPORTED A RIGHT SIDE DEFLATION. DEVICE REMAINS IMPLANTED.
THE DEVICE WAS EXPLANTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 413999 | STYLE 68 SALINE FILLED BREAST IMPLANT | PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE | FWM | ALLERGAN (COSTA RICA) | 1411348 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 63 YR | Required Intervention |