FDA Adverse Event Injury Summary report: N

STYLE 68 SALINE FILLED BREAST IMPLANT

MDR report key: 8619843 · Received May 17, 2019

Report

Report Number
9617229-2019-03261
Event Type
Injury
Date Received
May 17, 2019
Date of Event
April 17, 2019
Report Date
September 4, 2019
Manufacturer
ALLERGAN (COSTA RICA)
Product Code
FWM
PMA / PMN Number
P990074
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
WA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

ANALYSIS OF THE RETURNED DEVICE IDENTIFIED: WEAR ABRASION, CREASE FOLD, OPENING ON POSTERIOR BROWN AND YELLOW PARTICLES DEVICE INNER SURFACE AND RED PARTICLES IN FILL CHANNEL. LEAK TEST AND MICROSCOPIC ANALYSIS WAS PERFORMED WHICH IDENTIFIED: OPENING CREASE SHARP, OPENING CREASE SMOOTH AND OBSERVED VOID >1 MM IN NON-TEXTURED, VALVE-TO SHELL-BOND AREA. THE FILL TEST INSPECTION WAS PERFORMED, THE RESULT IS NO BLOCKAGE. BASED ON THE DEVICE ANALYSIS THE FINAL ASSESSMENT IS: AN OPENING CREASE SHARP ON ANTERIOR DUE TO FOLD FLAW OPENING, AND A SMOOTH CREASE OPENING ON POSTERIOR DUE TO FOLD FLAW OPENING.

Additional Manufacturer Narrative · 1

A REVIEW OF THE DEVICE HISTORY RECORD HAS BEEN COMPLETED. NO DEVIATIONS OR NON-CONFORMANCES NOTED. FURTHER INFORMATION FROM THE REPORTER REGARDING EVENT, PRODUCT, OR PATIENT DETAILS HAS BEEN REQUESTED. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME. THE REASON FOR REOPERATION IS DEFLATION. THIS IS A KNOWN POTENTIAL ADVERSE EVENT ADDRESSED IN THE PRODUCT LABELING.

Description of Event or Problem · 1

HEALTHCARE PROFESSIONAL REPORTED A RIGHT SIDE DEFLATION. THE DEVICE WAS EXPLANTED.

Description of Event or Problem · 1

HEALTHCARE PROFESSIONAL REPORTED A RIGHT SIDE DEFLATION. DEVICE REMAINS IMPLANTED.

Description of Event or Problem · 1

THE DEVICE WAS EXPLANTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
413999 STYLE 68 SALINE FILLED BREAST IMPLANT PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE FWM ALLERGAN (COSTA RICA) 1411348

Patients

Seq Age Sex Outcome Treatment
1 63 YR Required Intervention