FDA Adverse Event
Malfunction
Summary report: N
HIP GENERAL OFFSET-30° BROACH HANDLE - RIGHT
MDR report key: 8619673
·
Received May 17, 2019
Report
- Report Number
- 3005180920-2019-00388
- Event Type
- Malfunction
- Date Received
- May 17, 2019
- Date of Event
- April 17, 2019
- Report Date
- May 17, 2019
- Manufacturer
- MEDACTA INTERNATIONAL SA
- Product Code
- LXH
- UDI-DI
- 07630040742115
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- 003
Narratives
Additional Manufacturer Narrative · 1
BATCH REVIEW PERFORMED ON 14 MAY 2019: LOT 1854515: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 14-DEC-2018. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, TWO OTHER SIMILAR EVENTS HAVE BEEN REPORTED ON THIS SAME LOT ((B)(4)). VISUAL INSPECTION PERFORMED BY R&D PROJECT MANAGER: THE SPRING BALL PLUNGER IS CAME APART. IT IS POSSIBLE THIS OCCURRENCE WAS INFLUENCED BY VARIATION IN PRODUCTION/ASSEMBLY HOWEVER THIS CANNOT BE CONFIRMED.
Description of Event or Problem · 1
DURING A SURGERY THE LOCKING BALL OF THE INSTRUMENT FELL OUT OF THE HANDLE. THE BALL DID NOT FALL IN THE PATIENT. THE STRAIGHT AMIS-BROACH HANDLE HAS BEEN USED TO COMPLETE THE SURGERY SUCCESSFULLY, NO CRITICAL DELAY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 415258 | HIP GENERAL OFFSET-30° BROACH HANDLE - RIGHT | SURGICAL INSTRUMENT FOR HIP SURGERY | LXH | MEDACTA INTERNATIONAL SA | 1854515 | 07630040742115 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |