ORBERA365 INTRAGASTRIC BALLOON SYSTEM
Report
- Report Number
- 3006722112-2019-00082
- Event Type
- Injury
- Date Received
- May 15, 2019
- Date of Event
- September 12, 2018
- Report Date
- October 1, 2018
- Manufacturer
- APOLLO ENDOSURGERY, INC.
- Product Code
- LTI
- PMA / PMN Number
- P140008
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- PHYSICIAN
Narratives
"ADDITIONAL INFORMATION REGARDING THIS LATE MEDWATCH REPORT: DURING A FEBRUARY 2019 INTERNAL REVIEW OF POST-MARKET SURVEILLANCE DATA FOR ALL APOLLO PRODUCTS, APOLLO EXECUTIVE MANAGEMENT WAS MADE AWARE THAT COMPLAINTS ASSOCIATED WITH THE CE MARKED ORBERA365 (12-MONTH INTRAGASTRIC BALLOON) WERE NOT BEING ASSESSED FOR REPORTABILITY CONSISTENT WITH THE REQUIREMENTS OF 21 CFR 803. CONSIDERATION FOR REPORTABILITY IN THE US WAS INCORRECTLY ESTABLISHED BASED ON THE DIFFERENCE IN THE INDICATIONS FOR USE (I.E. 12-MONTH PLACEMENT VS 6-MONTH PLACEMENT) VS CONSIDERATIONS FOR SIMILAR DEVICES MARKETED BY APOLLO. THE ORBERA365 12-MONTH INTRAGASTRIC BALLOON IS A SIMILAR DEVICE AS COMPARED TO THE ORBERA 6-MONTH BALLOON APPROVED VIA PMA P140008, AND AS SUCH, COMPLAINTS WILL BE ASSESSED FOR MDR REPORTABILITY GOING FORWARD." THE REPORTER OF THE EVENT WAS ASKED TO RETURN THE PRODUCT FOR ANALYSIS, HOWEVER IT WAS INDICATED THAT IT WOULD NOT BE RETURNED. DEVICE LABELING ADDRESSES THE REPORTED EVENTS AS FOLLOWS: WARNINGS AND PRECAUTIONS: THE PHYSIOLOGICAL RESPONSE OF THE PATIENT TO THE PRESENCE OF THE ORBERA365¿ SYSTEM BALLOON MAY VARY DEPENDING UPON THE PATIENT'S GENERAL CONDITION AND THE LEVEL AND TYPE OF ACTIVITY. THE TYPES AND FREQUENCY OF ADMINISTRATION OF DRUGS OR DIET SUPPLEMENTS AND THE OVERALL DIET OF THE PATIENT MAY ALSO AFFECT THE RESPONSE. EACH PATIENT MUST BE MONITORED CLOSELY DURING THE ENTIRE TERM OF TREATMENT IN ORDER TO DETECT THE DEVELOPMENT OF POSSIBLE COMPLICATIONS. EACH PATIENT SHOULD BE INSTRUCTED REGARDING SYMPTOMS OF DEFLATION, GASTROINTESTINAL OBSTRUCTION, ULCERATION AND OTHER COMPLICATIONS WHICH MIGHT OCCUR, AND SHOULD BE ADVISED TO CONTACT HIS/HER PHYSICIAN IMMEDIATELY UPON THE ONSET OF SUCH SYMPTOMS. POSSIBLE COMPLICATIONS OF THE USE OF THE ORBERA365¿ SYSTEM INCLUDE: INJURY TO THE DIGESTIVE TRACT DURING PLACEMENT OF THE BALLOON IN AN IMPROPER LOCATION SUCH AS IN THE ESOPHAGUS OR DUODENUM. THIS COULD CAUSE BLEEDING OR EVEN PERFORATION, WHICH COULD REQUIRE A SURGICAL CORRECTION FOR CONTROL. ABDOMINAL OR BACK PAIN, EITHER STEADY OR CYCLIC. INJURY TO THE LINING OF THE DIGESTIVE TRACT AS A RESULT OF DIRECT CONTACT WITH THE BALLOON, GRASPING FORCEPS, OR AS A RESULT OF INCREASED ACID PRODUCTION BY THE STOMACH. THIS COULD LEAD TO ULCER FORMATION WITH PAIN, BLEEDING OR EVEN PERFORATION. SURGERY COULD BE NECESSARY TO CORRECT THIS CONDITION.
REPORTED AS: A PATIENT WITH THE ORBERA365 INTRAGASTRIC BALLOON HAD ACUTE ABDOMINAL PAIN. PERITONITIS. PATIENT SENT TO URGENCIES AT THE HOSPITAL. THE SCANNER SHOWS A GASTRIC PERFORATION. THE BALLOON WAS EXPLANTED, THE GASTRIC PERFORATION WAS STITCHED AND DEEP CLEANING PERFORMED. ADDITIONAL INFORMATION NOTED, 1 CM PERFORATION WAS LOCATED IN THE ANTRUM.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 404898 | ORBERA365 INTRAGASTRIC BALLOON SYSTEM | INTRAGASTRIC BALLOON | LTI | APOLLO ENDOSURGERY, INC. | B-50012 | AF01881 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 57 YR | Required Intervention |