MI-1000 DUAL CEILING
Report
- Report Number
- 2028295-2018-00003
- Event Type
- Malfunction
- Date Received
- May 14, 2019
- Manufacturer
- MEDICAL ILLUMINATION INTRNATION INC.
- Product Code
- FST
- PMA / PMN Number
- K100388
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
WE RECEIVED THE RETURNED LIGHT UNDER CUSTOMER COMPLAINT : (B)(4) AND AN INVESTIGATION WAS INITIATED. THE UNIT WAS VISUALLY INSPECTED AND THE ARM SUPPORT FITTING (P/N 1002036) WAS FOUND TO HAVE FAILED. NO OTHER DAMAGE WAS OBSERVED. HAVING SEE A SIMILAR FAILURE WITHIN THE PAST SIX MONTHS, WE PERFORMED MULTIPLE CUT CROSS-SECTION ANALYSIS TO RULE OUT POROSITY AS A CAUSE OF FAILURE. THE MATERIAL POROSITY LOOKED NORMAL AND NO STRESS-RELATED FACTORS WERE FOUND. THE PRODUCT COULD NOT BE CAPACITY TESTED DUE TO THE FAILURE LOCATION SO IT WAS REBUILT AND TESTED WITH A P/N 1002036. WE THEN TESTED THE SYSTEM TO THE FAILURE ANALYSIS OF THE MI-1000 SYSTEM DATED 07/16/2015. THE UNIT PASSED ALL TESTS. WE COULD NOT DETERMINE THE ULTIMATE CAUSE OF THE FAILURE OF THE P/N 1002036. THE LOAD TEST WAS PERFORMED AT 6X AND THE UNIT SHOWED ON DEGRADATION. NOTE: THIS REPORT IS LATE BECAUSE IT WAS SENT TO THE TEST EMDR INSTEAD OF PRODUCTION.
ON (B)(6) 2017 AN INCIDENT WAS REPORTED TO MEDICAL ILLUMINATION REGARDING AN MI-1000 ARM SUPPORT FITTING CAUSING THE LIGHT TO FALL. THERE WERE NO INJURIES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 403487 | MI-1000 DUAL CEILING | MINOR SURGICAL LIGHT | FST | MEDICAL ILLUMINATION INTRNATION INC. | 061515-V |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |