FDA Adverse Event Malfunction Summary report: N

OPTILENE 7/0 (0,5) 45CM DSMP11 (M) RCP

MDR report key: 8606144 · Received May 13, 2019

Report

Report Number
3003639970-2019-00408
Event Type
Malfunction
Date Received
May 13, 2019
Date of Event
April 15, 2019
Report Date
May 13, 2019
Manufacturer
B.BRAUN SURGICAL SA
Product Code
GAW
PMA / PMN Number
SEE H10
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

AESCULAP INC. (IMPORTER) IS SUBMITTING THIS REPORT ON BEHALF OF B. BRAUN SURGICAL S.A. (MANUFACTURER). EXEMPTION NUMBER: E2014012. REPORTED DEVICE NOT MARKETED IN THE U.S., HOWEVER, SIMILAR DEVICES OR DEVICES THAT SHARE COMPONENTS, RAW MATERIALS, PROCESS METHODS OR OTHER TECHNOLOGICAL CHARACTERISTICS ARE REGISTERED WITHIN THE U.S. K133890. INVESTIGATION: SAMPLES RECEIVED: 4 UNOPENED RACE-PACKS. ANALYSIS AND RESULTS: THERE ARE NO PREVIOUS COMPLAINTS OF THIS CODE-BATCH. WE MANUFACTURED 396 UNITS OF THIS CODE-BATCH. THERE ARE 108 UNITS IN OUR STOCK. WE HAVE RECEIVED FOUR CLOSED SAMPLES FOR ANALYSIS. WE HAVE TESTED THE KNOT PULL TENSILE STRENGTH OF THE SAMPLES RECEIVED AND THE RESULTS FULFIL THE REQUIREMENTS OF THE EUROPEAN PHARMACOPOEIA (EP):0.21 KGF IN AVERAGE AND 0.208 KGF IN MINIMUM (EP REQUIREMENTS: 0.10 KGF IN AVERAGE AND 0.036 KGF IN MINIMUM). REVIEWED THE BATCH MANUFACTURING RECORD, THIS PRODUCT HAD A NORMAL PROCESS AND THE RESULTS DURING THE PROCESS FULFIL USP/EP AND B.BRAUN SURGICAL REQUIREMENTS. REMARKS: WHEN WORKING WITH OPTILENE® SUTURE MATERIAL, GREAT CARE SHOULD BE TAKEN TO AVOID ANY CRUSHING OR CRIMPING DAMAGE OF THE MONOFILAMENT BY INSTRUMENTS SUCH AS FORCEPS OR NEEDLE HOLDERS.

Description of Event or Problem · 0

IT WAS REPORTED THE THREAD BREAKS VERY EASY INTRAOPERATIVELY. DURING AN EYELID LIFT/EYELID PROCEDURE, THE THREAD BREAKS VERY EASY. NO PATIENT HARM. NO OTHER INFORMATION HAS BEEN PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
397912 OPTILENE 7/0 (0,5) 45CM DSMP11 (M) RCP OTHER SUTURE GAW B.BRAUN SURGICAL SA C3090606 118161V004

Patients

Seq Age Sex Outcome Treatment
1