FDA Adverse Event Malfunction Summary report: N

BD DISCARDIT II 2 ML SYRINGE

MDR report key: 8601163 · Received May 10, 2019

Report

Report Number
3002682307-2019-00297
Event Type
Malfunction
Date Received
May 10, 2019
Date of Event
April 25, 2019
Report Date
June 7, 2019
Manufacturer
BECTON DICKINSON, S.A.
Product Code
FMF
PMA / PMN Number
N/A
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION: BD HAS BEEN PROVIDED WITH THE AFFECTED SAMPLE FOR CATALOG 300928 LOT 1811147 TO INVESTIGATE FOR THIS RECORD. AFTER THE EVALUATION OF THE RETURNED SAMPLE, THE FOREIGN MATTER WAS IDENTIFIED AS EMBEDDED CONTAMINATION IN THE PLUNGER. AS A RESULT, BD WAS ABLE TO VERIFY THE REPORTED ISSUE. THE EMBEDDED PARTICLES DEFECT WAS PRODUCED IN THE SCREW OF THE INJECTION MOLDING MACHINE. DUE TO THE HIGH WORKING TEMPERATURES (ABOUT 300ºC), SOME PARTICLES OF PLASTIC AND REST OF OIL FROM THE POLYETHYLENE CAN REMAIN STUCK ON THE INTERNAL WALLS OF THE MOLD AND GET BURNT. DURING NORMAL PRODUCTION SOME PARTICLES MAY BE DETACHED FROM THE SCREW BEING INJECTED AND REMAINING EMBEDDED IN MOLDED PIECE. THIS A COSMETIC DEFECT WHICH HAS NO RISK TO THE HEALTH BECAUSE THE PLASTIC PIECE IS EMBEDDED IN THE NEEDLE HUB WITHOUT POSSIBILITY OF BEING DETACHED FROM IT. DHR SHOWED NO INDICATION OF THE ALLEGED DEFECT.

Description of Event or Problem · 0

IT WAS REPORTED THAT THERE WERE GREY STAINS, SIMILAR TO MOLD WITH A BD DISCARDIT¿ II 2 ML SYRINGE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM GERMAN TO ENGLISH: THE SYRINGE PLUNGER HAS BLACK GREY STAINS, SIMILAR TO MOULD AT THE NEW, NOT USED PRODUCT IN THE UNDAMAGED PACKAGING.

Additional Manufacturer Narrative · 1

INITIAL REPORTER PHONE #: UNKNOWN. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THERE WERE GREY STAINS, SIMILAR TO MOLD WITH A BD DISCARDIT¿ II 2 ML SYRINGE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM (B)(6) TO ENGLISH: THE SYRINGE PLUNGER HAS BLACK GREY STAINS, SIMILAR TO MOULD AT THE NEW, NOT USED PRODUCT IN THE UNDAMAGED PACKAGING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
391782 BD DISCARDIT II 2 ML SYRINGE SYRINGE FMF BECTON DICKINSON, S.A. 1811147

Patients

Seq Age Sex Outcome Treatment
1 Other