FDA Adverse Event
Malfunction
Summary report: N
GE OEC 9900
MDR report key: 859747
·
Received March 23, 2007
Report
- Report Number
- 1720753-2007-00305
- Event Type
- Malfunction
- Date Received
- March 23, 2007
- Date of Event
- February 27, 2007
- Report Date
- March 23, 2007
- Manufacturer
- GE OEC MEDICAL SYSTEMS INC.
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
A GE SERVICE REPRESENTATIVE PERFORMED AN ON SITE INVESTIGATION. THE FAILURE COULD NOT BE DUPLICATED. THE GIE, FFB BOARDS, AND PS1 POWER SUPPLY WERE REPLACED DURING THE SERVICE CALL. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE COLLIMATOR ON THE 9900 SYSTEM CLOSED DOWN DURING THE CASE. NO PATIENT INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | GE OEC 9900 | FLUROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS INC. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 68 YR |