FDA Adverse Event Malfunction Summary report: N

BD¿ IV SET SN404 W/O BP Y-CONN

MDR report key: 8595096 · Received May 9, 2019

Report

Report Number
2243072-2019-00885
Event Type
Malfunction
Date Received
May 9, 2019
Date of Event
April 24, 2019
Report Date
June 12, 2019
Manufacturer
BECTON DICKINSON
Product Code
FOZ
PMA / PMN Number
N/A
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KS
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION: ONE SAMPLE WAS RETURNED TO SBDM, LOT NUMBER IS UNKNOWN. CLAMPING TEST UNDER NORMAL USING CONDITION: SBDM CONDUCT CLAMPING TEST UNDER NORMAL USING CONDITION FOR THE COMPLAINT SAMPLE, THE CLAMP COULD HOLD MEDICINE AND THERE WAS NO DRUG INFUSION. CLAMPING TEST UNDER HIGH AIR PRESSURE: SBDM CONDUCT CLAMPING TEST UNDER AIR PRESSURE(0.51MPA) FOR THE COMPLAINT SAMPLES, THE CLAMP COULD HOLD AIR AND THERE WAS NO AIR BUBBLE LEAKAGE. HOWEVER, SBDM FOUND LEAKAGE IN THE ANOTHER RECEIVED COMPLAINT SAMPLE OF (B)(4) WHICH WAS THE SAME COMPLAINT ISSUE. SBDM THUS CONDUCTED FURTHER INVESTIGATION ABOUT TUBE INNER & OUTER DIAMETER MEASUREMENT AND ECCENTRICITY OF ROLLER COMPONENT FOR COMPLAINT SAMPLE AND HOUSE SAMPLE. TUBE INNER & OUTER DIAMETER MEASUREMENT: USING PROFILE PROJECTOR, SBDM MEASURED THE INNER AND OUTER DIAMETER OF BOTH COMPLAINT SAMPLE AND HOUSE SAMPLE, THE DIAMETER ARE WITHIN SPECIFICATION. ROLLER ECCENTRICITY TEST (UNIT: MM): USING VERNIER CALIPER, SBDM MEASURED THE LENGTH FROM CENTER TO EDGE OF ROLLER FOR BOTH RECEIVED COMPLAINT SAMPLE (CAVITY NO.: 11, MINIMUM LENGTH: 8.22. MAXIMUM LENGTH: 8.35) & HOUSE SAMPLE, CONCLUSION WAS THERE SEEMS TO ECCENTRICITY IN THE ROLLER. HOUSE SAMPLE INSPECTION: SBDM INSPECTED 30 PCS FROM POTENTIAL LOTS 2811132, 2811152 AND 2811232, NO ABNORMALITY WAS OBSERVED. SBDM ALSO MEASURED THE ROLLER CLAMP CONCENTRICITY, ALL 30 ROLLER CLAMPS WAS WITHIN SPECIFICATIONS. DEVICE HISTORY RECORD REVIEW: SBDM REVIEWED THE MANUFACTURING RECORD FOR LOTS 2811132, 2811152 AND 2811232, NO ABNORMALITY WAS OBSERVED. ROOT CAUSE: FROM INVESTIGATIONS, SBDM CONDUCTED INSPECTION OF THE COMPLAINT SAMPLE & HOUSE SAMPLES FOR LEAKAGE TEST UNDER NORMAL CONDITION AND HIGH PRESSURE. NO LEAKAGE WAS OBSERVED ON THE RECEIVED COMPLAINT SAMPLE. SBDM ALSO CONDUCTED FURTHER TEST OF TUBE INNER & OUTER MEASUREMENT AND CHECK ECCENTRICITY OF ROLLER COMPONENT. THIS IS BECAUSE EVEN THERE WAS NO LEAKAGE IN THIS COMPLAINT SAMPLE, SBDM FOUND LEAKAGE ON SOME OF THE RECEIVED SAMPLE IN OTHER CASE (B)(4), PRODUCT: IV SET AN120 W/O BP. SBDM FOUND THERE WAS ECCENTRICITY IN THE ROLLER COMPONENT (THE LENGTH FROM CENTER TO EDGE OF ROLLER (MIN: 8.21MM, MAX: 8.4MM & GAP: 0.19MM) IN THE OTHER CASE ((B)(4)). SBDM THEN CHECKED ROLLER LENGTH FROM CENTER TO EDGE FOR 21 DIFFERENT LOT FROM RETENTION SAMPLES (TOTAL: 210EA) AND THE MAXIMUM GAP WAS 0.19 AND AVERAGE GAP WAS 0.09. IN CONCLUSION, SBDM ASSUMED THAT WHEN NARROW LOCATION OF THE ROLLER WHICH HAS ECCENTRICITY AND THE SMALLEST SIZE OF TUBE ARE MET WHEN ROLLER CLAMP IS IN LOCKED POSITION, MEDICINE LEAKAGE MIGHT OCCURRED INFREQUENTLY.

Description of Event or Problem · 0

IT WAS REPORTED THAT BD¿ IV SET SN404 W/O BP Y-CONN TUBE WAS CLAMPED AND THE DRUG WAS INFUSED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: AFTER CLAMPING THE TUBE, DRUG WAS INFUSED.

Additional Manufacturer Narrative · 1

DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. THE MANUFACTURING LOCATION FOR THIS PRODUCT IS GREATER ASIA. THIS SITE IS AN OEM MANUFACTURING SITE. (B)(4). MEDICAL DEVICE EXPIRATION DATE: UNKNOWN. DEVICE EVALUATED BY MFR: A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED, BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED. DEVICE MANUFACTURE DATE: UNKNOWN.

Description of Event or Problem · 1

IT WAS REPORTED THAT BD¿ IV SET SN404 W/O BP Y-CONN TUBE WAS CLAMPED AND THE DRUG WAS INFUSED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: AFTER CLAMPING THE TUBE, DRUG WAS INFUSED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
387124 BD¿ IV SET SN404 W/O BP Y-CONN INTRAVASCULAR CATHETER FOZ BECTON DICKINSON UNKNOWN

Patients

Seq Age Sex Outcome Treatment
1 Other