FDA Adverse Event Malfunction Summary report: N

METRIX AIR

MDR report key: 8584676 · Received May 6, 2019

Report

Report Number
1000113657-2019-00458
Event Type
Malfunction
Date Received
May 6, 2019
Date of Event
April 11, 2019
Report Date
May 6, 2019
Manufacturer
TRIVIDIA HEALTH, INC.
Product Code
NBW
UDI-DI
00021292007720
PMA / PMN Number
K150052
Report Source
Manufacturer report
Reporter Location
SC
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(MANUFACTURER NARRATIVE = T, CORRECTED DATA = F) INTERNAL REPORT # (B)(4). RETURNED METER EVALUATED WITH NO DEFECT FOUND. TEST STRIPS NOT RETURNED FOR EVALUATION. MOST LIKELY UNDERLYING ROOT CAUSE MLC-20 USER'S TEST STRIP HAD POOR STORAGE. TEST STRIP UDI#: (B)(4). NOTE: MANUFACTURER CONTACTED CUSTOMER IN A FOLLOW-UP CALL TO ENSURE THAT THE REPLACEMENT PRODUCTS RESOLVED THE INITIAL CONCERN - ABLE TO ESTABLISH CONTACT WITH CUSTOMER WHO INDICATED THAT METER IS WORKING AS DESIGNED.

Description of Event or Problem · 1

CONSUMER REPORTED COMPLAINT FOR HIGH AND ERRATIC BLOOD GLUCOSE TEST RESULTS. THE CUSTOMER IS CONCERNED WITH TESTS RESULTS FROM RESULTS OBTAINED OF 155, 256, 150, 162 AND 165 MG/DL. RESULTS REPORTED WERE NOT PERFORMED BACK TO BACK. THE CUSTOMER'S EXPECTED FASTING BLOOD GLUCOSE TEST RESULT RANGE IS 110 - 125 MG/DL. THE CUSTOMER FEELS WELL AND DID NOT REPORT ANY SYMPTOMS. MEDICAL ATTENTION IS NOT REPORTED AS A RESULT OF THE ACTUAL BLOOD GLUCOSE RESULTS. DURING THE CALL A BACK TO BACK BLOOD TEST WAS NOT PERFORMED BY THE CUSTOMER. THE PRODUCT IS NOT STORED ACCORDING TO SPECIFICATION AND IS STORED IN THE KITCHEN. THE TEST STRIP LOT MANUFACTURER'S EXPIRATION DATE IS 08/31/2020 AND OPEN VIAL DATE IS 04/2019. THE METER MEMORY WAS REVIEWED FOR PREVIOUS TEST RESULT HISTORY: (B)(6).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
376386 METRIX AIR BLOOD GLUCOSE SYSTEM NBW TRIVIDIA HEALTH, INC. METRIX AIR MV3194S 00021292007720

Patients

Seq Age Sex Outcome Treatment
1 0 YR