PROG VALVE INLINE W SG
Report
- Report Number
- 1226348-2019-00131
- Event Type
- Malfunction
- Date Received
- May 6, 2019
- Date of Event
- April 18, 2019
- Report Date
- April 23, 2019
- Manufacturer
- RAYNHAM
- Product Code
- JXG
- PMA / PMN Number
- K041296
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PHYSICIAN
Narratives
UDI INFORMATION (B)(4). SAMPLE WAS RECEIVED FOR EVALUATION. THE VALVE WAS VISUALLY INSPECTED; A CUT/TEAR IN THE SILICONE HOUSING IN THE NEEDLE CHAMBER WAS NOTED. THE POSITION OF THE CAM WHEN VALVE WAS RECEIVED WAS 140MMH2O. THE TECHNICIAN WAS ABLE TO CONFIRM THE PROBLEM REPORTED BY THE CUSTOMER. THE ROOT CAUSE FOR THE CUT/TEAR SILICONE HOUSING IS DUE TO A POINTED OR SHARP OBJECT COMING INTO CONTACT WITH THE SILICONE HOUSING. AS NOTED IN THE IFU SILICONE HAS A LOW TEAR/CUT RESISTANCE. MANUFACTURING RECORDS WERE REVIEWED AND FOUND NO ANOMALIES. BASED ON THE RESULTS OF THE INVESTIGATION, THE REPORTED ISSUE IS CONFIRMED. TRENDS WILL BE MONITORED FOR THIS AND SIMILAR ISSUES. AT PRESENT, WE CONSIDER THIS COMPLAINT TO BE CLOSED.
N/A.
TO DATE THE DEVICE INVOLVED IN THE REPORTED INCIDENT HAS NOT BEEN RECEIVED FOR EVALUATION. AN INVESTIGATION HAS BEEN INITIATED BASED ON THE REPORTED INFORMATION.
A SURGEON REPORTED THAT THE VALVE (ID 823162-PROG VALVE INLINE W SG) WAS PREPPED FOR USE ON (B)(6) 2019. IT WAS CONFIRMED THAT THE VALVE HAD NO PROBLEM WHEN THE AIR WAS REMOVED. HOWEVER WHEN THE VALVE WAS ABOUT TO BE IMPLANTED, THE SALINE WAS LEAKED FROM THE RESERVOIR. THEREFORE THE SURGEON REPLACED IT WITH A NEW VALVE AND THERE WAS NO ISSUE CONFIRMED. THE SURGEON REQUESTED AN INVESTIGATION REPORT. THERE WAS NO SURGICAL DELAY AND THERE WAS NO ADVERSE CONSEQUENCE TO THE PATIENT. NO FURTHER INFORMATION WAS PROVIDED BY HOSPITAL.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 376907 | PROG VALVE INLINE W SG | CHPV WITH SG | JXG | RAYNHAM | 3240361 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |