LIFEVEST WCD 4000 SYSTEM
Report
- Report Number
- 3008642652-2019-03110
- Event Type
- Death
- Date Received
- May 3, 2019
- Date of Event
- April 6, 2019
- Report Date
- April 24, 2019
- Manufacturer
- ZOLL MANUFACTURING CORPORATION
- Product Code
- MVK
- UDI-DI
- 00855778005005
- PMA / PMN Number
- P010030
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- OTHER
Narratives
DEVICE EVALUATION OF MONITOR SN (B)(4) AND ELECTRODE BELT SN (B)(4) HAS BEEN COMPLETED. DURING THE INCOMING FUNCTIONAL TESTING, A 1HZ SIMULATED NORMAL SINUS RHYTHM SIGNAL WAS APPLIED TO THE ECG ELECTRODES, FOLLOWED BY A 5HZ SIMULATED TREATABLE ARRHYTHMIA SIGNAL WHICH VERIFIED PROPER PERFORMANCE OF THE DETECTION ALGORITHM. DURING THE TRANSITION TO THE 5HZ SIGNAL, THE DEVICE WAS CONFIRMED TO PROPERLY ENTER INTO A TREATMENT SEQUENCE WHICH INCLUDES A VERIFICATION OF THE TACTILE VIBRATION ALARM, AUDIO MESSAGING, AND SIREN ALARMS, AS WELL AS A TEST OF THE PULSE DELIVERY CIRCUITRY. THE PULSE DELIVERY CIRCUITRY TEST VERIFIED PROPER CHARGING OF THE HIGH VOLTAGE CAPACITORS AND PROPER DELIVERY OF FIVE FULL ENERGY 150J BIPHASIC PULSES. THE FUNCTIONAL TESTING CONFIRMED PROPER RESPONSE BUTTON FUNCTIONALITY, ECG ACQUISITION, DETECTION ALGORITHM PERFORMANCE, AND PULSE DELIVERY FUNCTIONALITY. THERE IS NO INDICATION OF A PRODUCT MALFUNCTION. DEVICE MANUFACTURE DATES: MONITOR: 09/20/2017. ELECTRODE BELT: 08/25/2011.
A US DISTRIBUTOR CONTACTED ZOLL TO REPORT THAT A PATIENT WAS TREATED AND PASSED AWAY ON (B)(6) 2019. IT WAS REPORTED THAT THE PATIENT HAD BEEN WEARING THE LIFEVEST UNTIL BEING ADMITTED TO THE ICU. THE DEVICE WAS REMOVED IN THE ICU BECAUSE RESUSCITATION EFFORTS WERE BEING ATTEMPTED. IT WAS REPORTED THAT PRIOR TO BEING ADMITTED TO THE ICU, THE PATIENT WAS IN A DIFFERENT HOSPITAL FLOOR AND HAD BEEN CODING THAT MORNING AND HAD LOST CONSCIOUSNESS. THE NURSE DID NOT RECALL HEARING THE LIFEVEST ALARM OR GIVE COMMANDS. PER REVIEW OF THE EVENT, THE PATIENT WAS TREATED ONCE BY THE LIFEVEST PRIOR TO PASSING. THE DEVICE INITIALLY DETECTED AN ARRHYTHMIA AT 09:14:13 ON (B)(6) 2019 WHILE THE PATIENT WAS IN AN IDIOVENTRICULAR RHYTHM AT 65 BPM WITH PVCS AND CPR ARTIFACT. THE DEVICE EXITED THE DETECTION SEQUENCE AT 09:14:48. THERAPY ELECTRODE PLACEMENT FAULT FLAGS ARE SEEN FROM 09:26:35 TO 10:31:37. BETWEEN 10:56:34 AND 10:57:33, THE DEVICE DETECTED AN ARRHYTHMIA TWICE WHILE THE PATIENT WAS IN BRADYCARDIA AT 50 BPM WITH CPR ARTIFACT. THE DETECTION SEQUENCE WAS EXITED AT 10:57:36. AT 10:57:59 THE DEVICE DETECTED AN ARRHYTHMIA WHILE THE PATIENT WAS IN BRADYCARDIA AT 25 BPM WITH CPR ARTIFACT. GEL WAS RELEASED AT 10:58:24 AND AT 10:58:43, THE DEVICE DELIVERED A LOW ENERGY TREATMENT SHOCK WHILE THE PATIENT WAS IN BRADYCARDIA AT 25 BPM WITH CPR ARTIFACT. THE POST-SHOCK RHYTHM WAS BRADYCARDIA AT 50 BPM WITH PVCS. PER THE CONTINUOUS ECG RECORDINGS, FROM 10:58:43 UNTIL THE ELECTRODE BELT WAS DISCONNECTED AT 11:35:22 ON (B)(6) 2019, THE PATIENT WAS IN BRADYCARDIA FROM 40 BPM TO 50 BPM WITH CPR ARTIFACT. THE RHYTHM THEN TRANSITIONED TO SINUS TACHYCARDIA AT 110 BPM DEGRADING TO ASYSTOLE WITH MOTION ARTIFACT. CPR ARTIFACT CONTRIBUTED TO THE FALSE DETECTION. THE RESPONSE BUTTONS WERE NOT PRESSED DURING THE EVENT. DURING THE DELIVERY OF THE SHOCK, AN IMPEDANCE OF 0 OHMS WAS OBSERVED AND THE ENERGY DELIVERED IN THE SHOCK WAS 5 JOULES. THE LOW ENERGY SHOCK WAS LIKELY DUE TO THE PULSE BEING DELIVERED INTO AN OPEN LOAD. THIS IS CONSISTENT WITH 0 OHM IMPEDANCE. THE CAUSE OF THE OPEN LOAD AND REPORTED IMPEDANCE WAS LIKELY DUE TO THE THERAPY PADS ON THE LIFEVEST NOT MAKING FULL CONTACT WITH THE PATIENT AT THE TIME OF THE DEFIBRILLATION EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 370524 | LIFEVEST WCD 4000 SYSTEM | WEARABLE CARDIOVERTER DEFIBRILLATOR | MVK | ZOLL MANUFACTURING CORPORATION | WCD 4000 | 00855778005005 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 62 YR | Death| O |