STEM: QUADRA-H CEMENTLESS, HA COATED STD STEM SIZE 6
Report
- Report Number
- 3005180920-2019-00340
- Event Type
- Injury
- Date Received
- May 3, 2019
- Date of Event
- April 3, 2019
- Report Date
- June 26, 2019
- Manufacturer
- MEDACTA INTERNATIONAL SA
- Product Code
- JDI
- UDI-DI
- 07630030802072
- PMA / PMN Number
- K082792
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AS
- Reporter Occupation
- 003
Narratives
VISUAL INSPECTION PERFORMED BY THE R&D PROJECT MANAGER ON THE STEM RECEIVED: FEMORAL STEM SHOWS HA LAYER STILL INTACT ON THE PROXIMAL PART OF THE STEM. SIGNS DUE TO REVISION ON THE NECK. IN THE EVENT DESCRIPTION HAS BEEN ADDED THAT ALSO THE HEAD HAS BEEN REVISED.
ABOUT 2 YEARS AND 9 MONTHS AFTER PRIMARY THE SURGEON REVISED THE PATIENT HIP FOR A SUSPECTED INFECTION. THE STEM COME OUT EASY FROM FEMUR DURING THE REVISION. THE STEM AND THE HEAD WERE REVISED. THE SURGERY WAS COMPLETED SUCCESSFULLY.
BATCH REVIEW PERFORMED ON 29 APR 2019: LOT 156215: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 15-JAN-2016. EXPIRATION DATE: 2021-01-02. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THIS SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY OTHER SIMILAR REPORTED EVENT. OTHER DEVICES WERE INVOLVED IN THE EVENT: BALL HEADS: MECTACER 01.29.214 BIOLOX DELTA CERAMIC BALL HEAD 12/14 Ø 40 SIZE L + 4 LOT. 150463 (K112115): (B)(4) ITEMS MANUFACTURED AND RELEASED ON 10-APR-2015. EXPIRATION DATE: 2020-02-29. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THIS SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY OTHER SIMILAR REPORTED EVENT. CUP: MPACT 01.32.158DH ACETABULAR SHELL Ø58 TWO-HOLES LOT. 162911 (K132879): (B)(4) ITEMS MANUFACTURED AND RELEASED ON 22-JUN-2016. EXPIRATION DATE: 2021-06-01. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, ALL THE ITEMS OF THIS SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY OTHER SIMILAR REPORTED EVENT. LINER: MPACT 01.32.4048HCT FLAT PE HC LINER Ø40/F LOT. 157212 (K122641): (B)(4) ITEMS MANUFACTURED AND RELEASED ON 05-FEB-2016. EXPIRATION DATE: 2021-01-19. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THIS SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY OTHER SIMILAR REPORTED EVENT. CLINICAL EVALUATION PERFORMED BY MEDICAL AFFAIRS DIRECTOR: FEMORAL COMPONENT REVISION SURGERY OCCURRED 2 YEARS AND 7 MONTHS AFTER CEMENTLESS TOTAL HIP ARTHROPLASTY IN A HEAVY (B)(6) MAN. RADIOGRAPHIC IMAGES PROVIDED SHOW THE PRESENCE OF OSTEOLYTIC ZONES AROUND THE STEM AND BONE QUALITY ALTERATIONS IN THE GREATER TROCHANTER REGION. ACCORDING TO THE REPORT AND THE RADIOLOGICAL FINDINGS, AN INFECTION COULD BE A POSSIBLE CAUSE OF THIS EVENT. HOWEVER, ON THE BASIS OF THE INFORMATION RECEIVED, THIS HYPOTHESIS CANNOT BE CONFIRMED.
ABOUT 2 YEARS AND 9 MONTHS AFTER PRIMARY THE SURGEON REVISED THE PATIENT HIP FOR A SUSPECTED INFECTION. THE STEM COME OUT EASY FROM FEMUR DURING THE REVISION. THE STEM WAS REVISED. THE SURGERY WAS COMPLETED SUCCESSFULLY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 370002 | STEM: QUADRA-H CEMENTLESS, HA COATED STD STEM SIZE 6 | HIP CEMENTLESS STEM | JDI | MEDACTA INTERNATIONAL SA | 156215 | 07630030802072 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 69 YR | Required Intervention |