FDA Adverse Event Malfunction Summary report: N

HEATER-COOLER SYSTEM 3T

MDR report key: 8577146 · Received May 3, 2019

Report

Report Number
9611109-2019-00306
Event Type
Malfunction
Date Received
May 3, 2019
Date of Event
April 5, 2019
Report Date
May 3, 2019
Manufacturer
LIVANOVA DEUTSCHLAND
Product Code
DWC
UDI-DI
04033817901099
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THERE WAS NO PATIENT INVOLVEMENT. THE HEATER-COOLER 16-02-80 IS NOT DISTRIBUTED IN THE USA, BUT IT IS SIMILAR TO HEATER-COOLER 16-02-85, WHICH IS DISTRIBUTED IN THE USA (510(K) NUMBER: K052601). LIVANOVA (B)(4) MANUFACTURES THE HEATER-COOLER SYSTEM 3T. THE INCIDENT OCCURRED IN (B)(6). THE FAULT WAS DUE TO THE DAMAGED COMPRESSOR OF THE COOLING SYSTEM. THE DEVICE WAS REMOVED FROM SERVICE. A REVIEW OF THE DHR COULD NOT IDENTIFY ANY DEVIATIONS OR NONCONFORMITIES RELEVANT TO THE ISSUE. CORRECTIVE ACTIONS ARE IN PROGRESS FOR THIS ISSUE.

Description of Event or Problem · 1

LIVANOVA (B)(4) RECEIVED A REPORT THAT A HEATER-COOLER SYSTEM 3T TRIGGERED THE FUSES DURING PRIMING. THERE WAS NO PATIENT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
370828 HEATER-COOLER SYSTEM 3T CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS DWC LIVANOVA DEUTSCHLAND 16-02-80 04033817901099

Patients

Seq Age Sex Outcome Treatment
1