FDA Adverse Event Malfunction Summary report: N

DCI EQUIPMENT TRACK MOUNT LIGHT ASSEMBLY

MDR report key: 857384 · Received May 30, 2007

Report

Report Number
3004517290-2007-00001
Event Type
Malfunction
Date Received
May 30, 2007
Date of Event
May 1, 2007
Report Date
May 1, 2007
Manufacturer
DENTAL EQUIPMENT LLC (DBA NEWBERG MANUFACTURING)
Product Code
EAZ
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

AFTER THE MFR, DENTAL EQUIPMENT LLC, REC'D THE CUSTOMER COMPLAINT, AN INTERNAL INVESTIGATION REVEALED THAT THE DENTIST REC'D FOUR DUAL TRACK LIGHTS CONSISTING OF FOUR TRACK ASSEMBLIES FROM PELTON & CRANE (P&C) AND EIGHT LIGHT ASSEMBLIES FROM DENTAL EQUIPMENT LLC. AFTER EVALUATING THE ELECTRICAL SPECS FROM THE P&C TRANSFORMER, IT WAS DISCOVERED THAT DENTAL EQUIPMENT LLC'S LIGHT ASSEMBLIES WERE NOT PROPERLY RATED TO HANDLE THE 2 .9 VAC SUPPLIED FROM THE P&C TRANSFORMER. THE MFR'S LIGHT ASSEMBLY WAS EXCEEDED BY 7 VAC. AS A RESULT OF THE HIGHER VOLTAGE, THE LIGHT BULB OPERATED AT A HIGHER TEMPERATURE CAUSING THE LIGHT SHIELD TO MELT THROUGH THE PLASTIC LIGHT COVER. ALL EIGHT DENTAL EQUIPMENT LLC LIGHT ASSEMBLIES WERE REPLACED WITH EIGHT P&C LIGHT ASSEMBLIES PER P&C FACTORY SPEC.

Description of Event or Problem · 1

WHILE THE DENTIST WAS USING THE DENTAL LIGHT DURING PT TREATMENT, THE DENTIST NOTICED THAT THE LIGHT SHIELD WAS MELTING. AS THE DENTIST MOVED THE LIGHT BULB SHIELD ASSEMBLY AWAY FROM THE TREATMENT AREA, IT DROPPED ON HIS GLOVED HAND AND CAME IN CONTACT WITH A TOWEL LOCATED AROUND THE PT'S CHIN AREA. THERE WERE NO INJURES REPORTED. THE BULB SHIELD WEIGHS 12 OUNCES AND WAS APPROX 27 INCHES AWAY DURING USE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 DCI EQUIPMENT TRACK MOUNT LIGHT ASSEMBLY DENTAL OPERATING LIGHT EAZ DENTAL EQUIPMENT LLC (DBA NEWBERG MANUFACTURING) 1232-110 NA

Patients

Seq Age Sex Outcome Treatment
1 YR Required Intervention