CUP: MPACT ACETABULAR SHELL Ø52 NO-HOLE
Report
- Report Number
- 3005180920-2019-00330
- Event Type
- Injury
- Date Received
- May 2, 2019
- Date of Event
- April 2, 2019
- Report Date
- May 2, 2019
- Manufacturer
- MEDACTA INTERNATIONAL SA
- Product Code
- LPH
- UDI-DI
- 07630030810817
- PMA / PMN Number
- K132879
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- 003
Narratives
BATCH REVIEW PERFORMED ON 29 APR 2019: LOT 184599: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 19-OCT-2018. EXPIRATION DATE: 2023-10-10. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THIS SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY OTHER SIMILAR REPORTED EVENT. ADDITIONAL IMPLANTS INVOLVED: LINER: MPACT 01.32.3244HCT FLAT PE HC LINER Ø32/E (K103721), LOT 186965: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 09-OCT-2018. EXPIRATION DATE: 2023-09-24. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THIS SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY OTHER SIMILAR REPORTED EVENT. STEM: AMISTEM P 01.18.414 AMISTEM-P LAT STEM SIZE 4 (K173794), LOT 179949: 139 ITEMS MANUFACTURED AND RELEASED ON 03-MAY-2018. EXPIRATION DATE: 2023-04-23. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THIS SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY OTHER SIMILAR REPORTED EVENT. BALL HEADS: MECTACER 01.29.231H HEAD BIOLOX OPTION Ø 32 (K131518) [IMPLANTED ON (B)(6) 2019, DURING THE FIRST REVISION REPORTED IN THE MDR 2019-00290]. LOT 186574: (B)(4) ITEMS; MANUFACTURED AND RELEASED ON 12-NOV-2018. EXPIRATION DATE: 2023-10-28. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THIS SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY OTHER SIMILAR REPORTED EVENT. BALL HEADS: MECTACER 01.29.241A SLEEVE SIZE M (K131518) [IMPLANTED ON (B)(6) 2019, DURING THE FIRST REVISION REPORTED IN THE MDR 2019-00290]. LOT 183383: (B)(4); ITEMS MANUFACTURED AND RELEASED ON 25-JUL-2018. EXPIRATION DATE: 2023-07-16. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THIS SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY OTHER SIMILAR REPORTED EVENT.
REVISION SURGERY PERFORMED 13 DAYS AFTER THE PREVIOUS REVISION (PERFORMED WASHOUT AND HEAD REVISION ALSO DUE TO AN INFECTION, MDR 2019-00290) DUE TO AN INFECTION (THE PATHOGEN HAS BEEN IDENTIFIED AS STAPHYLOCOCCUS). THE SURGEON PERFORMED A WASHOUT AND REVISED THE STEM, HEAD, LINER, AND CUP. THE SURGERY WAS COMPLETED SUCCESSFULLY. THE PRIMARY CASE WAS PERFORMED ON (B)(6) 2019.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 367768 | CUP: MPACT ACETABULAR SHELL Ø52 NO-HOLE | CEMENTLESS ACETABULAR SHELL | LPH | MEDACTA INTERNATIONAL SA | 184599 | 07630030810817 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |