FDA Adverse Event Malfunction Summary report: N

2.7 DEGREE STRAIGHT SAGITAL SAW

MDR report key: 8572994 · Received May 2, 2019

Report

Report Number
3005985723-2019-00328
Event Type
Malfunction
Date Received
May 2, 2019
Date of Event
April 16, 2019
Report Date
July 8, 2019
Manufacturer
MAKO SURGICAL CORP.
Product Code
OLO
UDI-DI
00848486032111
PMA / PMN Number
K143752
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

REPORTED EVENT: STRAIGHT "SAW BLADE" ATTACHMENT STOPPED WORKING IN MID CUT. CASE TYPE: TKA. SURGICAL DELAY: 16-30 MINUTES. FUNCTIONAL INSPECTION: TURNING THE INPUT SHAFT BY BAND DOES NOT PRODUCE ANY MOTION AT THE BLADE PLATFORM. THE FAILURE MODE IS CONFIRMED. VISUAL INSPECTION: SHOWS THAT THE INPUT SHAFT BEARINGS HAVE BROKEN APART AND SLIPPED OFF THE SHAFT. SEE ATTACHED PICTURE. DIMENSIONAL INSPECTION: NOT PERFORMED AS THE VISUAL AND FUNCTIONAL INSPECTION ARE SUFFICIENT TO CONFIRM THE FAILURE. MATERIAL ANALYSIS: NOT PERFORMED AS NO MATERIAL FAILURE IS ALLEGED. PRODUCT HISTORY REVIEW: PRODUCT HISTORY REVIEW RESPECTIVELY SHOWS THE NUMBER OF DEVICES MANUFACTURED, DEVICES THAT FAILED INSPECTION, AND THEIR NC/NPR/QT NUMBERS IF APPLICABLE. DATE INSPECTED: 01-29-2019, DEVICES MANUFACTURED: (B)(4), DEVICES FAILED INSPECTION: 0, NC/NPR/QT NUMBERS: 0. PRODUCT HISTORY REVIEW SHOWS NO NON-CONFORMANCES WERE IDENTIFIED DURING INSPECTION. COMPLAINT HISTORY REVIEW: A REVIEW OF COMPLAINTS RELATED TO P/N: 212186, LOT NUMBER: 35030119 SHOWS NO ADDITIONAL COMPLAINT(S) RELATED TO THE FAILURE IN THIS INVESTIGATION. COMPLAINTS RELATED TO P/N: 212186 WILL BE TRACKED BY TREND REQUEST# (B)(4). CONCLUSION: THE COMPLAINT OF THE SAW ATTACHMENT NOT WORKING WAS CONFIRMED. THE FAILURE WAS TRACED TO THE BEARINGS ON THE INPUT SHAFT BREAKING APART. CORRECTIVE ACTION/PREVENTIVE ACTION: NO ACTION IS REQUIRED AT THIS TIME AS THERE IS NO INDICATION TO SUGGEST A PRODUCT NON-CONFORMITY OR UNANTICIPATED HAZARD.

Description of Event or Problem · 0

STRAIGHT "SAW BLADE" ATTACHMENT STOPPED WORKING IN MID CUT. CASE TYPE: TKA. SURGICAL DELAY: 16-30 MINUTES.

Additional Manufacturer Narrative · 1

¿AS PART OF NORMAL COMPLAINT FOLLOW-UP, AN EVALUATION OF THE EVENT HAS BEEN INITIATED BY MAKO SURGICAL. A SUPPLEMENTAL REPORT WILL BE SUBMITTED WHEN ADDITIONAL INFORMATION BECOMES AVAILABLE.¿

Description of Event or Problem · 1

STRAIGHT SAWBLADE ATTACHMENT STOPPED WORKING IN MID CUT. CASE TYPE: TKA. SURGICAL DELAY: 16-30 MINUTES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
368207 2.7 DEGREE STRAIGHT SAGITAL SAW STEREOTAXIC DEVICE, ROBOTICS OLO MAKO SURGICAL CORP. 35030119 / 3505311 00848486032111

Patients

Seq Age Sex Outcome Treatment
1 61 YR Hospitalization