FDA Adverse Event Other Summary report: N

CASTLE

MDR report key: 85721 · Received April 16, 1997

Report

Report Number
1314329-1997-90001
Event Type
Other
Date Received
April 16, 1997
Date of Event
April 8, 1997
Report Date
April 14, 1997
Manufacturer
GETINGE/CASTLE, INC.
Product Code
FLE
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

A SLOWLY LEAKING STEAM-TO-CHAMBER VALVE CAUSED THE CHAMBER TO PARTIALLY FILL WITH HOT WATER, WITH NO INCREASE IN PRESSURE IN THE CHAMBER. A WARNING LIGHT LABELED, "CAUTION DO NOT OPEN DOOR WHEN LIGHT IS ON" WAS ILLUMINATED, INDICATING PRESENCE OF WATER IN CHAMBER. THE OPERATOR IGNORED THE LIGHT AND OPENED THE DOOR. THE RESULTING FLOW OF HOT WATER CAUSED 2ND DEGREE BURNS. THE VALVE WAS DIAGNOSED AND REPAIRED BY A CO PRODUCT SUPPORT REP AND THE UNIT WAS TESTED TO CONFIRM THAT THE MACHINE WAS OPERATING PROPERLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CASTLE INSTRUMENT WASHER STERILIZER FLE GETINGE/CASTLE, INC. 3025 N/A

Patients

Seq Age Sex Outcome Treatment
1 UNKNOWN Required Intervention