FDA Adverse Event Malfunction Summary report: N

SYRINGE 10ML REG PR SALINE 10ML FILL

MDR report key: 8570502 · Received May 1, 2019

Report

Report Number
1911916-2019-00429
Event Type
Malfunction
Date Received
May 1, 2019
Date of Event
April 12, 2019
Report Date
May 17, 2019
Manufacturer
BD MEDICAL (BD WEST) MEDICAL SURGICAL
Product Code
NGT
UDI-DI
30382903065463
PMA / PMN Number
K161552
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WI, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION: ONE SAMPLE WAS RECEIVED. IT HAS RESIDUES OF DRIED BLOOD FROM THE LUER TO THE BOTTOM OF THE RUBBER STOPPER. THIS INDICATES THE SYRINGE WAS USED TO DRAW BLOOD. THE PLUNGER ROD- RUBBER STOPPER SEPARATION IS 3/32¿, THEREFORE FAILURE MODE IS VERIFIED. THE POSIFLUSH PRODUCT IS DESIGNED TO FLUSH THE LINES, NOT TO DRAWING BLOOD. A DEVICE HISTORY RECORD REVIEW WAS COMPLETED WITH ZERO DEFECTS FOUND. NO QUALITY NOTIFICATIONS WERE WRITTEN FOR THIS BATCH, NOR FOR THE ASSOCIATED ASSEMBLY BATCHES.

Description of Event or Problem · 0

IT WAS REPORTED THAT TWO SYRINGE 10ML REG PR SALINE 10ML FILL EXPERIENCED A DEFORMED STOPPER. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: MATERIAL NO.: 306546 BATCH NO.: 8172883. IT WAS REPORTED THAT THE PLUNGER SEPARATED FROM THE BLACK RUBBER STOPPER.

Additional Manufacturer Narrative · 1

A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED THAT TWO SYRINGE 10ML REG PR SALINE 10ML FILL EXPERIENCED A DEFORMED STOPPER. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: MATERIAL NO.: 306546, BATCH NO.: 8172883. IT WAS REPORTED THAT THE PLUNGER SEPARATED FROM THE BLACK RUBBER STOPPER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
361196 SYRINGE 10ML REG PR SALINE 10ML FILL SALINE, VASCULAR ACCESS FLUSH NGT BD MEDICAL (BD WEST) MEDICAL SURGICAL 8172883 30382903065463

Patients

Seq Age Sex Outcome Treatment
1 Other