ONCONTROL COAXIAL TRAY(11/13GA X 102/147
Report
- Report Number
- 3011137372-2019-00143
- Event Type
- Injury
- Date Received
- April 30, 2019
- Date of Event
- April 11, 2019
- Report Date
- April 24, 2019
- Manufacturer
- TELEFLEX MEDICAL
- Product Code
- KNW
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- PHYSICIAN
Narratives
QN# (B)(4). THE DHR FILE IS NOT AVAILABLE FOR REVIEW IN THE US. RECEIVED FOR INVESTIGATION ONE (1) 10CM BONE ACCESS NEEDLE SET (CANNULA ONLY), AND ONE (1) 14CM TREPHINE NEEDLE (BONE LESION BIOPSY NEEDLE) FOR INVESTIGATION. UPON PHYSICAL INSPECTION THE BENT 10CM BONE ACCESS NEEDLE WAS CONFIRMED. THE BEND IS LOCATED 1.5CM FROM THE DISTAL TIP END OF THE CANNULA. A REVIEW OF THE CERTIFICATE OF COMPLIANCE COULD NOT BE ACCOMPLISHED AS THE RECORDED LOT/BATCH IS NOT VALID. THIS REPORTED INCIDENT APPEARS TO BE DIRECTED MORE AT THE DETAILED DESCRIPTION OF THE INJURY SUFFERED BY THE PATIENT THAN WHAT MIGHT HAVE OCCURRED TO CAUSE THE BEND IN THE RETURNED CANNULA. THE INITIAL DESCRIPTION REFLECTS "A SNAGGING" FELT BY THE DR. WHEN TRYING TO RETRIEVE THE TREPHINE NEEDLE. THE REASON FOR THE "SNAGGING" FEEL WAS DUE TO A BENT BIOPSY CANNULA FROM THE NEEDLE SET, WHICH WAS RETURNED AS PART OF THE SAMPLES. THEN ANOTHER REFERENCE TO "A FRACTURE OF THE DISTAL END COAXIAL THAT REMAINED IN THE BODY...." DOES THIS DISCUSSION INDICATE A "FRACTURE" OF THE TREPHINE NEEDLE? THE TREPHINE NEEDLE WAS INSPECTED UNDER MAGNIFICATION AND THERE WAS NO "FRACTURE" FOUND. CLINICAL REGULATORY REVIEW STATES IN INTERNAL NOTE # 0006 REFERENCES, "THE PATIENT SUSTAINED A BONE FRACTURE". CONTINUING WITH INTERNAL NOTE #6, REFERENCE, "THE DOCTOR ONLY TILTED THE COAXIAL NEEDLE". TILTING A NEEDLE EMBEDDED IN A BONE WILL IN FACT BEND IT. OTHER REFERENCES GO ON TO MENTION FRAGMENTS OF THE CANNULA EMBEDDED IN THE PATIENT. QUESTIONS WHETHER SURGERY WAS REQUIRED TO REMOVE THE FRAGMENT, OR NOT. CONCLUSION: IF THERE IS A "FRACTURE OF THE DISTAL END COAXIAL", THE MISSING FRACTURE CANNOT BE IDENTIFIED ON THE SAMPLE. IF THERE IS A PIECE OF THE TREPHINE NEEDLE LEFT IN THE PATIENT, THEN WHY ALL THE DISCUSSION ABOUT THE FRACTURE OF THE COAXIAL? BASED ON THE CONVOLUTED DESCRIPTION OF THE EVENT AND THE PHYSICAL INSPECTION OF THE RETURNED COMPONENTS, THIS COMPLAINT CANNOT BE CONFIRMED.
IT WAS REPORTED THAT A TH4 VERTEBRAL BIOPSY WAS PERFORMED IN THE CONTEXT OF A HYPERMETABOLIC CONDENSING LESION CONTEXT OF PULMONARY AND PROSTATIC NEOPLASIA. TWO BONE BIOPSIES WERE PERFORMED ON THE UPPER SLOPE OF THE LESION. IT WAS ATTEMPTED TO PERFORM A THIRD BIOPSY AT THE LOWER SLOPE OF THE LESION: CAREFUL INCLINATION DOWNWARDS OF THE AXIAL AND BIOPSY TREPHINE, ADVANCING ALONG THE MARGIN WITH THE DRILL, SATISFACTORY TDM CONTROL. REMOVAL OF THE TREPHINE BIOPSY WITH THE DRIVER: AT THE TIME OF REMOVAL, THERE WAS THE SENSATION LIKE AN UNUSUAL SNAGGING FEELING. A XRAY WAS PERFORMED. THERE WAS A FRACTURE OF THE DISTAL END COAXIAL THAT REMAINED IN THE VERTEBRAL BODY AND THE RIGHT PEDICLE OF TH4 STRICT INTRAOSSEOUS POSITION.
(B)(4). THE DEVICE HAS NOT BEEN RETURNED FOR INVESTIGATION. TELEFLEX WILL CONTINUE TO MONITOR AND TREND RELATED EVENTS.
IT WAS REPORTED THAT A TH4 VERTEBRAL BIOPSY WAS PERFORMED IN THE CONTEXT OF A HYPERMETABOLIC CONDENSING LESION CONTEXT OF PULMONARY AND PROSTATIC NEOPLASIA. TWO BONE BIOPSIES WERE PERFORMED ON THE UPPER SLOPE OF THE LESION. IT WAS ATTEMPTED TO PERFORM A THIRD BIOPSY AT THE LOWER SLOPE OF THE LESION: CAREFUL INCLINATION DOWNWARDS OF THE AXIAL AND BIOPSY TREPHINE, ADVANCING ALONG THE MARGIN WITH THE DRILL, SATISFACTORY TDM CONTROL. REMOVAL OF THE TREPHINE BIOPSY WITH THE DRIVER: AT THE TIME OF REMOVAL, THERE WAS THE SENSATION LIKE AN UNUSUAL SNAGGING FEELING. A XRAY WAS PERFORMED. THERE WAS A FRACTURE OF THE DISTAL END COAXIAL THAT REMAINED IN THE VERTEBRAL BODY AND THE RIGHT PEDICLE OF TH4 STRICT INTRAOSSEOUS POSITION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 360457 | ONCONTROL COAXIAL TRAY(11/13GA X 102/147 | KNW | TELEFLEX MEDICAL | 102720 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |