CD1800 ANALYZER
Report
- Report Number
- 2919069-2007-00070
- Event Type
- Other
- Date Received
- May 25, 2007
- Date of Event
- April 26, 2007
- Report Date
- April 27, 2007
- Manufacturer
- ABBOTT LABORATORIES
- Product Code
- GKZ
- PMA / PMN Number
- K030513
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
Narratives
THE CUSTOMER STATED ALL REAGENT TUBING WAS WITHOUT OBSTRUCTION AND AIR BUBBLES, LYSE CAP INTACT, REAGENTS WERE CORRECT, AND ALL MAINTENANCE WAS UP TO DATE. ALL QC WAS WITHIN RANGE AND THE BUBBLE MIX IN THE PRE-MIX CUP AND BOTH TRANSDUCERS WAS VERIFIED. TEMPERATURE IN THE LABORATORY WAS CONSISTENT AT ALL TIMES. THE LAST CALIBRATION WAS PERFORMED IN 2007. THE DILUENT AND LYSE SYRINGES WERE WITHOUT POCKETS OR LEAKAGE; HOWEVER, THE 1 ML SAMPLE SYRINGE WAS LEAKING. THE CUSTOMER TECHNICAL ADVOCATE (CTA) RECOMMENDED REPLACING OR CLEANING THE SAMPLE SYRINGE, BUT THE CUSTOMER DECLINED TO PERFORM THE RECOMMENDED STEP. THE HGB REFERENCE = 1640 (TARGET IS 1800-2200). THE CTA INSTRUCTED THE CUSTOMER ON HOW TO PERFORM PRECISION. THE CTA DISPATCHED A FIELD SERVICE REPRESENTATIVE (FSR) TO FIX THE PROBLEM. THE FSR REPLACED THE LEAKING 1 ML SAMPLE SYRINGE AND TRAINED THE CUSTOMER ON HOW TO REPLACE THE PART. A REVIEW OF THE CELL DYN 1800 SYSTEM OPERATOR'S MANUAL, LIST NUMBER 07H80-01, REV. D, SECTION 9, SERVICE AND MAINTENANCE, AS-REQUIRED MAINTENANCE, FOUND GUIDELINES ON HOW TO CLEAN AND REPLACE THE 1 ML SYRINGE. IN ADDITION, THE COMPLAINT ANALYSIS AND TRENDING SYSTEM WAS REVIEWED AND NO ADVERSE TRENDS WERE IDENTIFIED. THIS IS THE FINAL REPORT. END OF REPORT.
THE CUSTOMER STATED THAT THE CELL DYN 1800 ANALYZER GENERATED A HEMOGLOBIN (HGB) RESULT OF 7.5 G/DL. THE SAMPLE WAS REPEATED 2 TIMES WITH RESULT OF 11.1 G/DL AND 10.9 G/DL. THE CUSTOMER CHECKED THE PATIENT'S PREVIOUS HGB RESULT AND IT WAS 10.9 G/DL. THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CD1800 ANALYZER | AUTOMATED HEMATOLOGY ANALYZER | GKZ | ABBOTT LABORATORIES | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK YR |