FDA Adverse Event Other Summary report: N

CD1800 ANALYZER

MDR report key: 856613 · Received May 25, 2007

Report

Report Number
2919069-2007-00070
Event Type
Other
Date Received
May 25, 2007
Date of Event
April 26, 2007
Report Date
April 27, 2007
Manufacturer
ABBOTT LABORATORIES
Product Code
GKZ
PMA / PMN Number
K030513
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE CUSTOMER STATED ALL REAGENT TUBING WAS WITHOUT OBSTRUCTION AND AIR BUBBLES, LYSE CAP INTACT, REAGENTS WERE CORRECT, AND ALL MAINTENANCE WAS UP TO DATE. ALL QC WAS WITHIN RANGE AND THE BUBBLE MIX IN THE PRE-MIX CUP AND BOTH TRANSDUCERS WAS VERIFIED. TEMPERATURE IN THE LABORATORY WAS CONSISTENT AT ALL TIMES. THE LAST CALIBRATION WAS PERFORMED IN 2007. THE DILUENT AND LYSE SYRINGES WERE WITHOUT POCKETS OR LEAKAGE; HOWEVER, THE 1 ML SAMPLE SYRINGE WAS LEAKING. THE CUSTOMER TECHNICAL ADVOCATE (CTA) RECOMMENDED REPLACING OR CLEANING THE SAMPLE SYRINGE, BUT THE CUSTOMER DECLINED TO PERFORM THE RECOMMENDED STEP. THE HGB REFERENCE = 1640 (TARGET IS 1800-2200). THE CTA INSTRUCTED THE CUSTOMER ON HOW TO PERFORM PRECISION. THE CTA DISPATCHED A FIELD SERVICE REPRESENTATIVE (FSR) TO FIX THE PROBLEM. THE FSR REPLACED THE LEAKING 1 ML SAMPLE SYRINGE AND TRAINED THE CUSTOMER ON HOW TO REPLACE THE PART. A REVIEW OF THE CELL DYN 1800 SYSTEM OPERATOR'S MANUAL, LIST NUMBER 07H80-01, REV. D, SECTION 9, SERVICE AND MAINTENANCE, AS-REQUIRED MAINTENANCE, FOUND GUIDELINES ON HOW TO CLEAN AND REPLACE THE 1 ML SYRINGE. IN ADDITION, THE COMPLAINT ANALYSIS AND TRENDING SYSTEM WAS REVIEWED AND NO ADVERSE TRENDS WERE IDENTIFIED. THIS IS THE FINAL REPORT. END OF REPORT.

Description of Event or Problem · 1

THE CUSTOMER STATED THAT THE CELL DYN 1800 ANALYZER GENERATED A HEMOGLOBIN (HGB) RESULT OF 7.5 G/DL. THE SAMPLE WAS REPEATED 2 TIMES WITH RESULT OF 11.1 G/DL AND 10.9 G/DL. THE CUSTOMER CHECKED THE PATIENT'S PREVIOUS HGB RESULT AND IT WAS 10.9 G/DL. THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CD1800 ANALYZER AUTOMATED HEMATOLOGY ANALYZER GKZ ABBOTT LABORATORIES NA NA

Patients

Seq Age Sex Outcome Treatment
1 UNK YR