FDA Adverse Event Malfunction Summary report: N

SERVO-U

MDR report key: 8564763 · Received April 30, 2019

Report

Report Number
8010042-2019-00289
Event Type
Malfunction
Date Received
April 30, 2019
Report Date
September 25, 2019
Manufacturer
MAQUET CRITICAL CARE AB
Product Code
CBK
PMA / PMN Number
K180098
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 0

THE REPORTED ISSUE WAS INVESTIGATED AT THE HOSPITAL BY OUR FIELD SERVICE ENGINEER (FSE). THE ISSUE COULD BE CONFIRMED, THE HUMIDIFIER/IV POLE HOLDER WAS LOOSE AND LEANING AWAY FROM THE MACHINE. NO PARTS HAVE BEEN SENT IN FOR INVESTIGATION. A SET SCREW HOLDS THE ALUMINUM PIECE TO THE SUPPORT ARM, THE SET SCREWS WERE STRIPPED OUT. THE ISSUE WAS RESOLVED BY REPLACING THE SCREWS AND SECURING THE POSITION OF THE HOLDER. A DESIGN IMPROVEMENT WAS IMPLEMENTED DURING (B)(6)2018 AND AN ADDITIONAL SET SCREW HAS BEEN INTRODUCED. OUR CONCLUSION INTO THIS MATTER IS THAT THE HUMIDIFIER HOLDER MOST LIKELY HAS BEEN EXPOSED TO A WEIGHT THAT IS GREATER THAN IT IS DESIGNED TO SUSTAIN.

Additional Manufacturer Narrative · 0

GETINGE USA SALES, LLC (IMPORTER) IS SUBMITTING THIS REPORT ON BEHALF OF MAQUET CRITICAL CARE AB (MANUFACTURER). REF. EXEMPTION #: E2018003. GETINGE USA SALES, LLC, 45 BARBOUR POND DRIVE, WAYNE, NJ 07470. CONTACT PERON: (B)(4).

Description of Event or Problem · 0

MANUFACTURER REFERENCE#:1915MCC1612. IMPORTER REFERENCE (B)(4).

Description of Event or Problem · 0

IT WAS REPORTED THAT THE SUPPORT ARM WAS COMING LOOSE AND WAS CAUSING THE ALUMINUM PIECE THAT CONNECTS TO THE SUPPORT RAIL TO PULL AWAY. THERE WAS NO PATIENT INVOLVEMENT. (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
359554 SERVO-U VENTILATOR, CONTINUOUS, FACILITY USE CBK MAQUET CRITICAL CARE AB

Patients

Seq Age Sex Outcome Treatment
1