FDA Adverse Event Injury Summary report: N

ERBE ICC 200 EA

MDR report key: 8564402 · Received April 30, 2019

Report

Report Number
9610614-2019-00012
Event Type
Injury
Date Received
April 30, 2019
Date of Event
February 26, 2019
Report Date
April 30, 2019
Manufacturer
ERBE ELEKTROMEDIZIN GMBH
Product Code
GEI
PMA / PMN Number
K933157
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE ESU WAS RETURNED AND THOROUGHLY INSPECTED/TESTED. A TECHNICAL SAFETY CHECK WAS PERFORMED ON THE GENERATOR. THIS INCLUDED AN ELECTRICAL SAFETY CHECK, A FUNCTION CHECK OF EACH OF THE EQUIPMENT'S FEATURES, AND A POWER OUTPUT CHECK. ALL FEATURES WERE/ARE FUNCTIONING PROPERLY WITHIN SPECIFICATIONS. IN CONCLUSION, NO ERBE EQUIPMENT PROBLEM WAS FOUND THAT WOULD HAVE CAUSED OR ATTRIBUTED TO THE EVENT. MOST LIKELY, THERE WERE MANY FACTORS INVOLVED IN THE REPORTED EVENT. HOWEVER, THE PATIENT'S CONDITION OF HAVING A BROAD BASED POLYP WAS A KEY FACTOR IN THE OUTCOME. SPECIFICALLY UPON THE INTERVENTION WORK, THE REMAINING TISSUE OF THE BOWL DID NOT STAY INTACT AS A RESULT OF THE NECROSIS AT THE ABLATION SITE AND THEN THE DELAYED PERFORATION. IN CONCLUSION, NO DETERMINATION COULD BE MADE AS TO THE CAUSE OF THE EVENT. NO TRENDS HAVE BEEN IDENTIFIED. ERBE USA, INC. IS NOW CLOSING THE FILE ON THIS EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE ELECTROSURGICAL UNIT WAS INVOLVED IN A PATIENT INCIDENT UPON A POLYPECTOMY. THE ESU WAS USED TO REMOVE A BROAD BASED POLYP. THE GENERATOR WAS USED WITH AN ERBE RETURN ELECTRODE CABLE (PART NUMBER 20194-058, LOT NUMBER NOT APPLICABLE) AND AN ERBE NESSY OMEGA RETURN ELECTRODE (PART NUMBER 20193-082, LOT NUMBER 160810-0816). THE RETURN ELECTRODE WAS PLACED ON THE PATIENT'S RIGHT THIGH. NO INFORMATION WAS PROVIDED REGARDING ANY OTHER ACCESSORY USED OR THE SPECIFIC SETTINGS EMPLOYED. DURING THE POLYPECTOMY, THERE HAD BEEN NO ABNORMALITIES OR DETECTABLE PERFORATION. HOWEVER DAYS LATER, THE PATIENT CAME BACK WITH ABDOMINAL PAIN. NECROSIS (INCLUDING A DELAYED PERFORATION) WAS FOUND AT THE ABLATION SITE OF THE POLYP. TO ADDRESS THE ISSUE, THE PATIENT UNDERWENT A PARTIAL RESECTION OF THE TRANSVERSE COLON WITH ADDITIONAL HOSPITALIZATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
358943 ERBE ICC 200 EA ELECTROSURGICAL UNIT GEI ERBE ELEKTROMEDIZIN GMBH ICC 200 EA

Patients

Seq Age Sex Outcome Treatment
1 64 YR Hospitalization| R