FDA Adverse Event Malfunction Summary report: Y

CATALYS SYSTEM

MDR report key: 8563493 · Received April 30, 2019

Report

Report Number
3005675890-2019-00040
Event Type
Malfunction
Date Received
April 30, 2019
Report Date
October 28, 2020
Manufacturer
JOHNSON & JOHNSON SURGICAL VISION, INC.
Product Code
OOE
UDI-DI
05050474609792
PMA / PMN Number
K121091
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

THIS CORRECTION MDR IS BEING SUBMITTED TO ADDRESS INACCURATE INFORMATION IN SECTION F AND G. 4 DATE ON THE XML VERSION OF THE PREVIOUS SUBMISSION CAUSED BY A SOFTWARE GLITCH IN THE TRANSMISSION PROCESS. THE INTERNAL NON-CONFORMANCE NUMBERS FOR THIS SOFTWARE ISSUE ARE NR-0148919 AND CAPA-010215.

Additional Manufacturer Narrative · 1

INVESTIGATION WAS COMPLETED FOR THE 4 REPORTED EVENTS. A REVIEW OF RECORDS RELATED TO THE DEVICE INCLUDING LABELING, COMPLAINT TRENDING, AND RISK DOCUMENTATION WAS PERFORMED. THERE WERE NO ISSUES FOUND WITH THE SYSTEM. ALL PERTINENT INFORMATION AVAILABLE TO JOHNSON & JOHNSON SURGICAL VISION, INC HAS BEEN SUBMITTED.

Additional Manufacturer Narrative · 1

SERIAL NUMBERS OF DEVICES: (B)(4). THE CATALYST SYSTEM MAY HAVE A REMOTE POTENTIAL EVENT WHERE LOSS OF SUCTION DURING LASER FIRING COULD CREATE A HAZARDOUS SITUATION THAT MAY RESULT IN SCORING OF THE POSTERIOR CORNEAL SURFACE. THEREFORE JOHNSON & JOHNSON SURGICAL VISION, INC HAS ISSUED AN ADVISORY NOTICE TO REINFORCE INSTRUCTIONS PROVIDED IN PRODUCT TRAINING AND IN THE OPERATOR¿S MANUAL REGARDING SUCTION LOSS DURING A PROCEDURE. ALL PERTINENT INFORMATION AVAILABLE TO JOHNSON & JOHNSON SURGICAL VISION, INC. THAT WAS RECEIVED DURING THE TIME PERIOD THAT COVERS THIS VOLUNTARY MALFUNCTION SUMMARY REPORT HAS BEEN SUBMITTED.

Description of Event or Problem · 1

THIS REPORT SUMMARIZES 7 MALFUNCTION EVENTS. THE EVENTS WERE RELATED TO SUCTION LOSS WHILE THE CATALYST LASER WAS FIRING. THERE WERE NO MEDICAL EVENTS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
358892 CATALYS SYSTEM CATALYS OOE JOHNSON & JOHNSON SURGICAL VISION, INC. CATALYS-U 05050474609792

Patients

Seq Age Sex Outcome Treatment
1