FDA Adverse Event Malfunction Summary report: N

BD ANGIOCATH¿ IV CATHETER

MDR report key: 8555815 · Received April 26, 2019

Report

Report Number
9610048-2019-00160
Event Type
Malfunction
Date Received
April 26, 2019
Date of Event
April 9, 2019
Report Date
June 11, 2019
Manufacturer
BECTON DICKINSON INDUSTRIAS CIRURGICAS, LTDA.
Product Code
FOZ
UDI-DI
30382903811121
PMA / PMN Number
K151698
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

H.6. INVESTIGATION: DHR REVIEW WAS DONE AND THIS LOT WAS REVIEWED REGARDING THE TESTS OF ¿FOREIGN MATTER/ VISIBLE SILICONE¿, WERE NOT EVIDENCED RECORDS THAT COULD LEAD TO THE CLAIMED DEFECT. WAS RECEIVED ON 24-APR-2019 IN DB SANDY ONE UNUSED UNIT WITHOUT PACKAGING FROM CATALOG NUMBER 381112, LOT NUMBER 7241831. ACCORDING TO THE VISUAL ANALYSIS OF THE SAMPLE CAN OBSERVE THAT THE FOREIGN MATTER FOUND IS SILICONE. IT SHOULD BE NOTED THAT THIS SILICONE DOES NOT CAUSE HARM TO THE WEARER AND IS USED TO AID IN THE SLIPPAGE OF THE CATHETER DURING VENIPUNCTURE. BASED ON THE INVESTIGATIONS CONDUCTED FOR COMPLAINTS OF SILICONE VISIBLE OF ANGIOCATH, IT HAS BEEN FOUND THAT THE ROOT CAUSE OF THIS ISSUE ON THE CATHETER IS CAUSED BY THE EXCESSIVE AMOUNT OF SILICONE THAT IS DISPENSED DURING THE SILICONIZATION OF THE CATHETER TIP TIPPING AND FINAL ASSEMBLY PROCESS. CAPA#450094 WAS INITIATED.

Description of Event or Problem · 0

IT WAS REPORTED THAT THERE WAS FOREIGN MATTER ON TIP OF NEEDLE WITH A BD ANGIOCATH¿ IV CATHETER. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM JAPANESE TO ENGLISH: FM IS ON THE NEEDLE TIP.

Additional Manufacturer Narrative · 1

(B)(6). A SAMPLE IS AVAILABLE FOR EVALUATION. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THERE WAS FOREIGN MATTER ON TIP OF NEEDLE WITH A BD ANGIOCATH¿ IV CATHETER. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM (B)(6) TO ENGLISH: FM IS ON THE NEEDLE TIP.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
347491 BD ANGIOCATH¿ IV CATHETER INTRAVASCULAR CATHETER FOZ BECTON DICKINSON INDUSTRIAS CIRURGICAS, LTDA. 7241831 30382903811121

Patients

Seq Age Sex Outcome Treatment
1 Other