FDA Adverse Event Malfunction Summary report: N

FUSION¿ ENT NAVIGATION SYSTEM

MDR report key: 8552029 · Received April 25, 2019

Report

Report Number
1723170-2019-01993
Event Type
Malfunction
Date Received
April 25, 2019
Date of Event
April 19, 2019
Report Date
May 22, 2019
Manufacturer
MEDTRONIC NAVIGATION, INC
Product Code
HAW
UDI-DI
00643169518452
PMA / PMN Number
K001284
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KS, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

A MEDTRONIC REPRESENTATIVE WENT TO THE SITE TO TEST THE EQUIPMENT. TESTING REVEALED THAT THE REPORTED ISSUE COULD NOT BE REPLICATED AND THE NAVIGATION SYSTEM FUNCTIONED AS DESIGNED. THE SYSTEM THEN PASSED THE SYSTEM CHECKOUT AND WAS FOUND TO BE FULLY FUNCTIONAL. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 1

A SOFTWARE ANALYSIS WAS INITIATED TO DETERMINE THE PROBABLE CAUSE. ANALYSIS FOUND ARCHIVE PRESENT IN FDR WAS REVIEWED AND CONFIRMED TO ONLY HAVE 5 SLICES IN THE DATA FOLDERS FOR THE EXAMS. THIS CORRELATES TO THE ISSUES FOUND BY (B)(4) WHEN ON SITE. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

MEDTRONIC RECEIVED INFORMATION THAT, WHEN LOADING A COMPUTED TOMOGRAPHY (CT) IMAGE ONTO THE NAVIGATION SYSTEM FROM A NEWTOM CGI EVO CT SCANNER, THE UNIVERSAL SERIAL BUS (USB) DEVICE USED COULD NOT BE LOADED ONTO THE NAVIGATION SYSTEM. A SECOND NAVIGATION SYSTEM AT THE SITE COULD LOAD THE EXAM WITHOUT ISSUE. A SECOND USB WAS BURNED WITHOUT RESOLUTION. THREE CDS WERE BURNED WITHOUT RESOLUTION. THE SURGEON OPTED TO COMPLETE THE PROCEDURE WITHOUT THE USE OF THE NAVIGATION SYSTEM. THERE WAS A REPORTED DELAY TO THE PROCEDURE OF MORE THAN 1 HOUR DUE TO THIS ISSUE. THERE WAS NO REPORTED IMPACT ON PATIENT OUTCOME.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
341768 FUSION¿ ENT NAVIGATION SYSTEM INSTRUMENT, STEREOTAXIC HAW MEDTRONIC NAVIGATION, INC 9733560XOM 00643169518452

Patients

Seq Age Sex Outcome Treatment
1 81 YR