HEATER-COOLER SYSTEM 3T
Report
- Report Number
- 9611109-2019-00281
- Event Type
- Malfunction
- Date Received
- April 25, 2019
- Report Date
- January 28, 2020
- Manufacturer
- LIVANOVA DEUTSCHLAND
- Product Code
- DWC
- Removal / Correction Number
- Z-2076/2081-2015
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
H.10: THE DEVICE WAS SENT TO LIVANOVA FOR DEEP DISINFECTION ON APRIL 16, 2019. ON MAY 21, 2019 LABORATORY TESTS CONDUCTED BEFORE DEEP DISINFECTION (PRE-DD) DID NOT CONFIRM THE REPORTED CONTAMINATION. AS PER THE INSTRUCTION FOR USE CP_IFU_16-XX-XX_ENG CHAPTER 6.3.3 MONITORING THE WATER QUALITY, THE TOTAL BACTERIA COUNT MUST NOT EXCEED THE LIMIT OF 100 CFU/ML. PRE-DD TESTS RESULTED TO BE WITHIN SPECIFICATIONS. THUS, THE REPORTED CONTAMINATION IS NOT CONFIRMED. BASED ON THE ABOVE MENTIONED EVIDENCES, IT IS LIKELY THAT A CONTAMINATION OF THE SAMPLES OCCURRED. AS PER THE INSTRUCTION FOR USE IT IS RECOMMENDED TO PAY ATTENTION TO HYGIENE MEASURE TO AVOID POTENTIAL CONTAMINATION OF THE SAMPLES.
SEE INITIAL.
THERE WAS NO KNOWN PATIENT INVOLVEMENT. PMA/510(K): THE HEATER-COOLER 16-02-80 IS NOT DISTRIBUTED IN THE USA, BUT IT IS SIMILAR TO HEATER-COOLER 16-02-85, WHICH IS DISTRIBUTED IN THE USA (510(K) NUMBER: K052601). RECALL NUMBER: LIVANOVA (B)(4) IMPLEMENTED A FIELD SAFETY NOTICE FOR DISINFECTION AND CLEANING OF HEATER-COOLER DEVICES. THE Z NUMBER IS Z-2076/2081-2015. LIVANOVA (B)(4) MANUFACTURES THE HEATER-COOLER SYSTEM 3T. THE INCIDENT OCCURRED IN (B)(6). THROUGH FOLLOW UP COMMUNICATION, LIVANOVA (B)(4) LEARNED THAT THE UNIT WAS CLEANED PER THE IFU BY ONE OF OUR TECHNICIANS. FURTHERMORE, THAT THE DEVICE IS PLACED INSIDE THE OPERATION THEATER, WITH THE FAN OF THE DEVICE POSITIONED OPPOSITE TO THE PATIENT, WITH A DISTANCE BETWEEN THE SURGERY FIELD AND THE DEVICE OF 2 METERS. THE DEVICE IS FULLY FUNCTIONAL. A REVIEW OF THE DHR COULD NOT IDENTIFY ANY DEVIATIONS OR NONCONFORMITIES RELEVANT TO THE ISSUE. CORRECTIVE ACTIONS ARE IN PROGRESS FOR THIS ISSUE. IF ANY ADDITIONAL INFORMATION PERTINENT TO THE REPORTED EVENT IS RECEIVED, IT WILL BE PROVIDED IN A SUPPLEMENTAL REPORT.
LIVANOVA (B)(4) RECEIVED A REPORT, THAT A HEATER-COOLER SYSTEM 3T WAS FOUND TO BE CONTAMINATED WITH PSEUDOMONAS AERUGINOSA AND OCHROBACTRUM ANTHROPI. THE LAB REPORT HAS BEEN SUPPLIED, TESTING PERFORMED AS CULTURE IN LIQUID MEDIUM. THERE IS NO KNOWN PATIENT INVOLVEMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 344951 | HEATER-COOLER SYSTEM 3T | CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS | DWC | LIVANOVA DEUTSCHLAND | 16-02-80 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|