Narratives
BATCH REVIEW PERFORMED ON 23 APRIL 2019: LOT 142339: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 03-JUN-2014. EXPIRATION DATE: 2019-05-31. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY SIMILAR REPORTED EVENT. CLINICAL EVALUATION PERFORMED BY MEDICAL AFFAIRS MANAGER: HIP REVISION SURGERY OCCURRED 4, 5 YEARS AFTER CEMENTLESS TOTAL HIP ARTHROPLASTY. RADIOGRAPHIC IMAGES PROVIDED SHOW THE PRESENCE OF RADIOLUCENT LINES AROUND THE STEM AND SIGNS OF STRESS SHIELDING.
THE PATIENT CAME IN, 4 YEARS AND 5 MONTHS AFTER PRIMARY SURGERY, COMPLAINING OF PAIN CAUSED BY A SUBSIDED STEM. THE SURGEON REVISED THE STEM AND HEAD WITH ANOTHER COMPANY'S PRODUCT AND THE SURGERY WAS COMPLETED SUCCESSFULLY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 345377 | STEM: AMISTEM H HA COATED STD STEM SIZE 1 | UNCEMENTED HIP STEM | LZO | MEDACTA INTERNATIONAL SA | 142339 | 07630030804076 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |