MICROSCAN PROMPT INOCULATION SYSTEM-D
Report
- Report Number
- 2919016-2006-00035
- Event Type
- Malfunction
- Date Received
- October 9, 2006
- Date of Event
- September 22, 2006
- Report Date
- September 22, 2006
- Manufacturer
- DADE BEHRING, INC.
- Product Code
- JSD
- Removal / Correction Number
- FCA 000264
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
PERFORMANCE TESTING OF CUSTOMER RETURNS AND RETENTION SAMPLES. IN-HOUSE TESTING PERFORMED AND CONFIRMED INSUFFICIENT GROWTH ISSUE WITH CUSTOMER PROVIDED SAMPLES AND ALSO WITH RETENTION SAMPLES. DADE BEHRING HAS INITIATED A FIELD CORRECTION FOR THIS ISSUE AND NOTIFIED CUSTOMER OF THE POTENTIAL FOR DISCREPANT MIC AND/OR IDENTIFICATION RESULTS WITH CLINICAL AND QC ISOLATES WITH THE IDENTIFIED PROMPT INOCULATION SYSTEM-D LOT.
ACCOUNT REPORTED TO DADE BEHRING THAT THEY HAVE OBSERVED QC OUT OF RANGE IDENTIFICATION RESULTS WITH ATCC 49147 (S. BOVIS) VS. IDX, BE, PGT BIOCHEMICALS AND ATCC 29213 (S. AUREUS) VS. IDX,URE,VP BIOCHEMICALS AND AMPICILLIN. ISSUE IS ONLY WITH QC. REPORT WAS ASSOCIATED WITH ONLY THE IDENTIFIED PROMPT LOT. NO PATIENT SAMPLES WERE AFFECTED. NO REPORT OF INJURY OR ILLNESS ASSOCIATED WITH THIS ISSUE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | MICROSCAN PROMPT INOCULATION SYSTEM-D | PANEL INOCULATING SUPPLIES | JSD | DADE BEHRING, INC. | NA | 20071127 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | YR |