MICROSCAN PROMPT INOCULATION SYSTEM-D
Report
- Report Number
- 2919016-2006-00040
- Event Type
- Malfunction
- Date Received
- October 9, 2006
- Date of Event
- September 29, 2006
- Report Date
- September 29, 2006
- Manufacturer
- DADE BEHRING, INC.
- Product Code
- JSD
- Removal / Correction Number
- FCA 000264
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KY, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
PERFORMANCE TESTING OF CUSTOMER RETURNS AND RETENTION SAMPLES. IN-HOUSE TESTING PERFORMED AND CONFIRMED INSUFFICIENT GROWTH ISSUE WITH CUSTOMER PROVIDED SAMPLES AND ALSO WITH RETENTION SAMPLES. DADE BEHRING HAS INITIATED A FIELD CORRECTION FOR THIS ISSUE AND NOTIFIED CUSTOMERS OF THE POTENTIAL FOR DISCREPANT MIC AND/OR IDENTIFICATION RESULTS WITH CLINICAL AND QC ISOLATES WITH THE IDENTIFIED PROMPT INOCULATION SYSTEM-D LOT.
ACCOUNT REPORTED TO DADE BEHRING THAT THEY HAD OBSERVED INSUFFICIENT GROWTH (NO RESULTS REPORTED) WITH SOME S. AUREUS AND OTHER STAPH. ISOLATES. ACCOUNT INDICATED THAT THEY DID NOT REPORT THE OBTAINED RESULTS. ISSUE WAS ONLY ASSOCIATED WITH THE IDENTIFIED PROMPT LOT. THERE WERE NO REPORTS OF INJURY OR ILLNESS ASSOCIATED WITH THIS ISSUE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | MICROSCAN PROMPT INOCULATION SYSTEM-D | PANEL INOCULATING SUPPLIES | JSD | DADE BEHRING, INC. | NA | 20071127 |