MIS 16CM STRAIGHT ATTACHMENT
Report
- Report Number
- 0001811755-2019-01491
- Event Type
- Malfunction
- Date Received
- April 24, 2019
- Date of Event
- January 1, 2019
- Report Date
- July 26, 2019
- Manufacturer
- STRYKER INSTRUMENTS-KALAMAZOO
- Product Code
- HBE
- UDI-DI
- 07613327096514
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- OTHER
Narratives
THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. SUPPLEMENTAL RATIONALE CORRECTED DATA:8 EVENTS WERE PREVIOUSLY REPORTED FOR THIS CATALOG NUMBER/DEVICE CODE COMBINATION DURING THE REPORTING PERIOD; HOWEVER, - 1 PREVIOUSLY REPORTED EVENT IN THIS REPORT SHOULD HAVE BEEN INCLUDED UNDER MFR REPORT #0001811755-2019-02229 - 7 PREVIOUSLY REPORTED EVENTS ARE INCLUDED IN THIS FOLLOW-UP RECORD. PRODUCT RETURN STATUS 7 DEVICES WERE RECEIVED. EVENT CONFIRMATION STATUS 1 REPORTED EVENT WAS CONFIRMED. 6 REPORTED EVENTS WERE NOT CONFIRMED. EVALUATION RESULTS 2 DEVICES WERE FOUND TO BE AFFECTED BY BROKEN DOWN LUBRICATION. 5 DEVICES HAD NO DEVICE PROBLEM FOUND.
THIS REPORT SUMMARIZES <NOE> 7 </NOE> MALFUNCTION EVENTS IN WHICH THE DEVICE REPORTEDLY OVERHEATED. 7 EVENTS HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT.
THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. REPORTED EVENTS: 8 EVENTS WERE REPORTED FOR THIS QUARTER. PRODUCT RETURN STATUS: 2 DEVICES WERE RECEIVED. 6 DEVICE INVESTIGATION TYPES HAVE NOT YET BEEN DETERMINED. EVENT CONFIRMATION STATUS: 2 DEVICE EVALUATIONS ARE STILL IN PROGRESS. ADDITIONAL INFORMATION: 8 DEVICES WERE NOT LABELED FOR SINGLE-USE. 8 DEVICES WERE NOT REPROCESSED AND REUSED.
THIS REPORT SUMMARIZES 8 MALFUNCTION EVENTS IN WHICH THE DEVICE REPORTEDLY OVERHEATED. 8 EVENTS HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 340708 | MIS 16CM STRAIGHT ATTACHMENT | DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED) | HBE | STRYKER INSTRUMENTS-KALAMAZOO | 07613327096514 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |