FDA Adverse Event Injury Summary report: N

LIFEPORT KIDNEY PUMP

MDR report key: 8545362 · Received April 23, 2019

Report

Report Number
MW5086095
Event Type
Injury
Date Received
April 23, 2019
Date of Event
April 10, 2019
Report Date
April 19, 2019
Manufacturer
ORGAN RECOVERY SYSTEMS, INC.
Product Code
KDN
Adverse Event
Yes
Report Source
Voluntary report
Reporter Location
MO, US
Reporter Occupation
003

Narratives

Description of Event or Problem · 1

LIFEPORT KIDNEY PUMP MALFUNCTION: THE RIGHT KIDNEY FOR ORGAN DONOR (B)(6) WAS PLACED ON THE LIFEPORT KIDNEY PUMP ON (B)(6) 2019. ON (B)(6) 2019 AT 1700 A RED FLASHING ERROR LIGHT WAS OBSERVED ON THE PUMP BUT NO ALARM SOUNDING. THE PUMP INFO WAS DOWNLOADED AND IT APPEARED THAT THE PUMP STOPPED PUMPING AT 1555. THE KIDNEY WAS TAKEN OFF THE PUMP, BUT WHEN THE CASSETTE WAS OPENED, THERE WAS NO VISIBLE FLUID IN THE CASSETTE. WE DO NOT KNOW HOW LONG THE KIDNEY WAS NOT IN COLD SOLUTION STORAGE AND COLD STORAGE SHOULD BE THE BACKUP MODE FOR ANY PUMP FAILURE. THEREFORE, THE KIDNEY WAS DECLINED FOR TRANSPLANT AND RESULTED IN A DISCARD OF THE ORGAN. ADD'L INFO: UPON FURTHER INVESTIGATION OF THE PUMP, THERE WAS ONLY ABOUT AN INCH OF MPS IN THE BOTTOM OF THE CASSETTE, THE REST OF THE SOLUTION WAS UNDER THE ICE BUCKET AND COMING OUT FROM THE BOTTOM OF THE MACHINE. ONCE WE IDENTIFIED THE ISSUE AND TOOK EVERYTHING APART, WE ALSO FLUSHED THE CASSETTES TO SEE IF THERE WERE ANY LEAKS. WE COULD NOT IDENTIFY ANY LEAKS IN THE CASSETTES. WE ALSO TIPPED THE MACHINE ON THE SIDE AND FOUND FLUID COMING OUT BETWEEN THE TOP AND BOTTOM OF THE PUMP. WE CALLED THE LIFEPORT PERFUSION HOTLINE AND THEY SENT A FORM FOR US TO FILL OUT AND INSTRUCTED UP TO SEND THE PUMP AND CASSETTE FOR FURTHER DEVICE INVESTIGATION. THE PUMP AND CASSETTES WERE SENT TO LIFEPORT FOR INSPECTION (B)(6) 2019. THIS EVENT WAS ALSO REPORTED TO (B)(6) AS A PT SAFETY EVENT (EVENT #(B)(6)) ON (B)(6) 2019. FDA SAFETY REPORT ID# (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
335827 LIFEPORT KIDNEY PUMP SYSTEM, PERFUSION, KIDNEY KDN ORGAN RECOVERY SYSTEMS, INC. LKT100 73588264
335828 CASSETTE SYSTEM, PERFUSION, KIDNEY KDN ORGAN RECOVERY SYSTEMS, INC. LKT200

Patients

Seq Age Sex Outcome Treatment
1 Other