FDA Adverse Event Malfunction Summary report: N

SUMMIT SWIVEL COLLIMATOR

MDR report key: 854510 · Received May 22, 2007

Report

Report Number
1450503-2007-00001
Event Type
Malfunction
Date Received
May 22, 2007
Date of Event
May 15, 2007
Report Date
May 22, 2007
Manufacturer
SUMMIT INDUSTRIES, INC.
Product Code
KPW
PMA / PMN Number
K874687
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

SUMMIT INDUSTRIES HAS NOT YET RECEIVED THE SUBJECT UNIT FOR EVALUATION. THE SUMMIT DEALER WAS NOTIFIED OF THE INCIDENT AND WENT TO THE USER FACILITY (UF) WHERE HE DETERMINED THAT THE COLLIMATOR SWIVEL RETAINING SCREWS HAD BECOME LOOSE AND/OR WERE MISSING WHEN THE COLLIMATOR FELL. ONE OR MORE OF THE REMAINING SCREWS HAD POSSIBLY PULLED OUT OF THE SWIVEL MOUNT ASSEMBLY, STRIPPING THE LAST TWO OF THE 8 THREADS IN THE MOUNTING ASSEMBLY. TO KEEP THE CLINIC OPERATING, HE DETERMINED THAT REPLACING THE ORIGINAL 1/4" RETAINING SCREWS WITH 1/2" ONES THAT FULLY ENGAGE THE REMAINING 6 THREADS AND ADDING LOCK-TIGHT WOULD BE ADEQUATE UNTIL A REPLACEMENT UNIT COULD ARRIVE FROM THE MFR, SUMMIT INDUSTRIES. AT THAT TIME, THE SUBJECT UNIT WILL BE RETURNED TO SUMMIT INDUSTRIES FOR EVALUATION.

Description of Event or Problem · 1

SUMMIT SWIVEL COLLIMATOR MODEL # D800 CAME FREE FROM THE TUBEMOUNT, LANDING ON THE TABLE TOP, BUMPING THE HAND OF A PT. THE PT WAS RECEIVING FOLLOW UP CARE AFTER WRIST SURGERY. SUBSEQUENT X-RAYS INDICATED THAT THE SURGICAL WORK WAS UNAFFECTED BY THE INCIDENT AND THAT NO ADD'L INJURY RESULTED. THE FAILURE APPEARS TO HAVE RESULTED FROM LOOSE OR MISSING HARDWARE ON THE SWIVEL ASSEMBLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SUMMIT SWIVEL COLLIMATOR BEAM LIMITING DEVICE KPW SUMMIT INDUSTRIES, INC. D800 ND513-1298

Patients

Seq Age Sex Outcome Treatment
1 85 YR Other