FDA Adverse Event Malfunction Summary report: N

INCISOR PLUS ELITE BLADE,3.5MM DSPL,DYO/

MDR report key: 8542065 · Received April 23, 2019

Report

Report Number
1219602-2019-00459
Event Type
Malfunction
Date Received
April 23, 2019
Date of Event
April 17, 2019
Report Date
June 6, 2019
Manufacturer
SMITH & NEPHEW, INC.
Product Code
HAB
UDI-DI
03596010555694
PMA / PMN Number
E527289
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ONE DYONICS DISPOSABLE 3.5MM INCISOR PLUS ELITE BLADE RETURNED. THE COMPLAINT STATED: ¿THE BLADE MADE A NOISE AND RELEASED SMOKE.¿ THE BLADE WAS FOUND SEIZED. ONCE THE INNER AND OUTER WERE FREED, WEAR BANDS AND SKIVING WAS VISIBLE ON THE INNER BLADE AND INSIDE THE OUTER BLADE CUP. ACCUMULATED SHAVINGS WERE STILL ATTACHED WHICH BOUND THE BLADE. DRAG WAS FELT WITH ROTATION OF THE INNER AND OUTER BLADES. THESE SYMPTOMS WERE CAUSED BY DISTAL PRESSURE APPLIED. AS A RESULT, BINDING, SHAVINGS AND SEIZING OCCURRED. NO ROOT CAUSE RELATED TO THE MANUFACTURE OF THE DEVICE WAS CONFIRMED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE BLADE MADE A NOISE AND RELEASED SMOKE. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITHOUT SIGNIFICANT DELAY USING A BACK-UP DEVICE. NO PATIENT INJURY OR OTHER COMPLICATIONS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
335556 INCISOR PLUS ELITE BLADE,3.5MM DSPL,DYO/ SAW, POWERED, AND ACCESSORIES HAB SMITH & NEPHEW, INC. 03596010555694