FDA Adverse Event Malfunction Summary report: N

I-STAT EC8+ CARTRIDGE

MDR report key: 8540828 · Received April 23, 2019

Report

Report Number
2245578-2019-00112
Event Type
Malfunction
Date Received
April 23, 2019
Date of Event
March 28, 2019
Report Date
March 24, 2020
Manufacturer
ABBOTT POINT OF CARE
Product Code
CHL
UDI-DI
10054749000071
PMA / PMN Number
K940918
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

APOC INCIDENT # (B)(4). THE INVESTIGATION WAS COMPLETED ON 03/20/2020. THE CUSTOMER REPORTED THAT ANALYZER S/N (B)(6) PRODUCED UNEXPECTED K AND BUN RESULTS WITH EC8+ CARTRIDGES FROM LOT K18343 AND UNEXPECTED PCO2 RESULTS WITH EC8+ CARTRIDGES FROM LOT H18235A. FAILURE ANALYSIS DID NOT REPRODUCE THE COMPLAINT. ANALYZER S/N (B)(6) FUNCTIONED ACCORDING TO SPECIFICATION THROUGHOUT TESTING AND PRODUCED RESULTS THAT WERE WITHIN THE ACCEPTANCE RANGES. A ROCKETWARE SEARCH SPANNING SIX MONTHS REVEALED TWO SIMILAR INCIDENTS (NON-REPRODUCIBLE UNEXPECTED PCO2 RESULTS) AND NO EVIDENCE OF A TREND. NO DEFICIENCY HAS BEEN IDENTIFIED.

Additional Manufacturer Narrative · 0

APOC INCIDENT # (B)(4). THE INVESTIGATION WAS COMPLETED ON 06/24/2019. THE CUSTOMER REPORTED THAT ANALYZER S/N (B)(4) PRODUCED UNEXPECTED K AND BUN RESULTS WITH EC8+ CARTRIDGES FROM LOT K18343 AND UNEXPECTED PCO2 RESULTS WITH EC8+ CARTRIDGES FROM LOT H18235A. FAILURE ANALYSIS COULD NOT BE PERFORMED AS ANALYZER S/N (B)(4) (APOC INCIDENT (B)(4)) HAS NOT BEEN RETURNED. A ROCKETWARE SEARCH SPANNING SIX MONTHS REVEALED NO SIMILAR INCIDENTS AND NO EVIDENCE OF A TREND. NO DEFICIENCY HAS BEEN IDENTIFIED.

Additional Manufacturer Narrative · 0

APOC INCIDENT #: (B)(4). THE INVESTIGATION WAS COMPLETED ON 06/12/2019. A REVIEW OF THE DEVICE HISTORY RECORD (DHR) CONFIRMED THE CARTRIDGE LOT MET FINISHED GOODS (FG) RELEASE CRITERIA. RETAINED TESTING MET THE ACCEPTANCE CRITERIA OUTLINED IN APPENDIX 1 OF Q04.01.003 REV. AD (PRODUCT COMPLAINT LEVEL 2 AND LEVEL 3 INVESTIGATION PROCEDURE). NO DEFICIENCY HAS BEEN DETERMINED FOR EC8+ LOT K18343.

Description of Event or Problem · 0

NA.

Additional Manufacturer Narrative · 1

APOC INCIDENT #: (B)(4). PRODUCT#: 6000-0000, SERIAL#: (B)(4), MFG DATE: 03/04/2010. ANALYZER BEING REQUESTED FOR INVESTIGATION. APOC LABELING WILL BE EVALUATED DURING THE INVESTIGATION AS PERTAINING TO THE EVENT.

Description of Event or Problem · 1

ON (B)(6) 2019, ABBOTT POINT OF CARE WAS CONTACTED BY A CUSTOMER REGARDING I-STAT EC8+ CARTRIDGES THAT YIELDED SUSPECTED DISCREPANT BUN AND POTASSIUM RESULTS ON A PATIENT. THERE WAS NO PATIENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT. METHOD: I-STAT, DATE: (B)(6) 2019, BUN: 124 MG, POTASSIUM: 79 MMOL/L, SAMPLE: A (SUSPECT INCORRECT RESULT. POSSIBLY 7.9). I-STAT, (B)(6) 2019, 18 MG, 4.4 MMOL/L, B. COLLECT AND TEST TIMES NOT PROVIDED. THERE ARE NO INJURIES ASSOCIATED WITH THIS EVENT. AT THIS TIME THERE IS NO REASON TO SUSPECT A MALFUNCTION EXISTS. THE REPORTING DECISION WAS BASED ON THE LIMITED INFORMATION AVAILABLE THAT SUGGESTS THE PRODUCT WAS NOT PERFORMING WITHIN THE VARIABILITY OF THE ASSAY. THERE WAS NO PATIENT INFORMATION. THE CUSTOMER STATES THAT THE POTASSIUM RESULT WAS 79 MMOL/L. HOWEVER, THE RESULT APPEARS TO BE INCORRECT. PER I-STAT SYSTEM MANUAL ART: 714174-00P, THE REPORTABLE RANGE FOR POTASSIUM IS 2.0 - 9.0 MM/L. THE CUSTOMER HAS REQUESTED THE ANALYZER BE REPLACED, AS THE ERRORS HAVE OCCURRED WITH TWO LOTS OF CARTRIDGES. THE ERRORS IN QUESTION HAVE NOT BEEN CONFIRMED AND THE PRODUCT HAS NOT BEEN REPLACED. THE INVESTIGATION IS UNDERWAY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
334385 I-STAT EC8+ CARTRIDGE EC8+ CARTRIDGE CHL ABBOTT POINT OF CARE NA K18343 10054749000071

Patients

Seq Age Sex Outcome Treatment
1