MDR report key: 8539386 · Received April 23, 2019

Report

Report Number
3002808157-2017-00879
Event Type
Injury
Date Received
April 23, 2019
Date of Event
August 26, 2014
Report Date
February 13, 2017
Manufacturer
SIEMENS HEALTHCARE GMBH
Product Code
LNH
PMA / PMN Number
K141977
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE ELECTRODES USED ARE LABELED MR CONDITIONAL BY THE MANUFACTURER; HOWEVER, THE USE OF THE ELECTRODES IS NOT APPROVED BY SIEMENS (UNILECT TM 4841P ELECTRODES, UNOMEDICAL LTD. (B)(4)). (B)(6).

Description of Event or Problem · 1

SIEMENS BECAME AWARE OF AN ADVERSE EVENT INVOLVING A SIEMENS MAGNETOM SKYRA SYSTEM THROUGH AN ARTICLE PUBLISHED IN THE JOURNAL OF RADIOLOGY AND IMAGING. AN UNUSUAL CASE OF A THIRD DEGREE SKIN BURN WAS REPORTED USING MRI ELECTROCARDIOGRAPHIC LEADS. THE ELECTRODES ARE LABELED MR CONDITIONAL BY THE MANUFACTURER, HOWEVER, THE ELECTRODES ARE NOT A SIEMENS APPROVED PRODUCT. THE VENDOR OF THE ELECTRODES WAS INFORMED BY THE USER WHEN THE INCIDENT OCCURRED, HOWEVER, SIEMENS WAS NOT INFORMED OF THE INCIDENT AS NO MALFUNCTION OF THE SIEMENS SYSTEM WAS OBSERVED. THE PATIENT WAS TREATED WITH AQUAGEL AND A FOAM BANDAGE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
333763 MAGNETOM SKYRA SYSTEM, NUCLEAR MAGNETIC RESONANCE IMAGING LNH SIEMENS HEALTHCARE GMBH 10432915

Patients

Seq Age Sex Outcome Treatment
1 74 YR Other