HEARTMATE II LEFT VENTRICULAR ASSIST SYSTEM
Report
- Report Number
- 2916596-2019-01982
- Event Type
- Injury
- Date Received
- April 22, 2019
- Date of Event
- August 9, 2018
- Report Date
- April 22, 2019
- Manufacturer
- THORATEC CORPORATION
- Product Code
- DSQ
- UDI-DI
- 00813024011170
- PMA / PMN Number
- P060040
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KY, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
APPROXIMATE AGE OF DEVICE- 6 YEARS, 8 MONTHS. MANUFACTURER'S INVESTIGATION CONCLUSION: UPON EVALUATION OF THE DEVICE, A THROMBUS WAS FOUND WITHIN THE DEVICE. DAMAGE TO THE BIONATE AND SILICONE JACKET LAYERS OF THE DRIVELINE WAS ALSO CONFIRMED. THE DEVICE WAS RETURNED ASSEMBLED WITH THE DRIVELINE (DL) SEVERED APPROXIMATELY 9.5¿ FROM THE PUMP HOUSING. A DISTAL SEGMENT WITH THE CONTROLLER CONNECTOR WAS ALSO RETURNED MEASURING APPROXIMATELY 27¿ IN LENGTH. THE SEALED INFLOW CONDUIT (INLET TUBE, FLEX SECTION, AND INLET ELBOW) WAS RETURNED ATTACHED TO THE PUMP¿S INLET PORT. THE APICAL SEWING RING WAS RETURNED SECURED TO THE INLET TUBE WITH GREEN SUTURES. THE OUTLET ELBOW WAS RETURNED ATTACHED TO THE PUMP¿S OUTLET PORT. THE SEALED OUTFLOW GRAFT AND SEALED OUTFLOW GRAFT BEND RELIEF WERE RETURNED DETACHED FROM THE OUTLET ELBOW. EXAMINATION OF THE OUTLET ELBOW REVEALED A PINK, TISSUE-LIKE DEPOSITION THAT APPEARED TO BE SLIGHTLY ADHERED TO THE TEXTURED SURFACE. THE THROMBUS FORMATION LACKED ANY EVIDENCE OF LAMINATION OR DENATURATION. ALTHOUGH A SPECIFIC CAUSE FOR THE OBSERVED THROMBUS FORMATION COULD NOT BE CONCLUSIVELY DETERMINED THROUGH THIS EVALUATION, IT DID NOT APPEAR TO AFFECT PUMP FUNCTION AS THE PATIENT WAS ROUTINELY TRANSPLANTED. THE REMAINING BLOOD-CONTACTING SURFACES REVEALED NO EVIDENCE OF DEVELOPED DEPOSITIONS OR THROMBUS FORMATIONS. THE DISTAL 24¿ OF DRIVELINE WAS WRAPPED EXTENSIVELY IN WHITE TAPE. UNDERNEATH THE TAPE THERE WAS A SIGNIFICANT AMOUNT OF DISCOLORATION AND SHIELD BREAKDOWN. A CLAMSHELL REPAIR (COMPLAINT #61746; CS-031263; WO #39810) WAS PRESENT WHICH WAS COVERING FATIGUE DAMAGE TO THE BIONATE LAYER AND METAL BRAIDED SHIELD APPROXIMATELY 0.5¿ FROM THE CONTROLLER CONNECTOR. THE DAMAGE TO THE BIONATE APPEARED TO RESULT FROM WEAR FATIGUE DUE TO REPETITIVE FLEXING. A SPECIFIC CAUSE FOR THE OBSERVED DAMAGE TO THE SILICONE JACKET COULD NOT BE DETERMINED. ELECTRICAL CONTINUITY TESTING OF THE REPAIRED DRIVELINE DID NOT REVEAL ANY DISCONTINUITIES OR SHORTS. THE PUMP WAS REASSEMBLED AND FUNCTIONALLY TESTED UNDER NORMAL OPERATING CONDITIONS USING A MOCK CIRCULATORY LOOP. THE DATA RETRIEVED FROM THAT TESTING REVEALED NORMAL PUMP POWER CONSUMPTION AND PRESSURE VALUES THAT MET MANUFACTURING SPECIFICATION AND THE DEVICE FUNCTIONED AS INTENDED. THE HEARTMATE II LVAS IFU LISTS DEVICE THROMBOSIS AS AN ADVERSE EVENT THAT MAY BE ASSOCIATED WITH THE USE OF HEARTMATE II LEFT VENTRICULAR ASSIST SYSTEM. SECTION 1 ¿INTRODUCTION¿ PROVIDES AN EXPLANATION OF EACH PUMP¿S PARAMETERS. SECTION 6 ¿PATIENT CARE AND MANAGEMENT¿ OUTLINES INDICATIONS OF PUMP THROMBOSIS AND HOW TO RESPOND TO SUCH EVENTS. SECTION 6 ALSO OUTLINES THE SUGGESTED ANTICOAGULATION THERAPY AND INR RANGE DURING USE OF THE HEARTMATE II LVAS. THE IFU ALSO PROVIDES INSTRUCTIONS FOR DEVICE IMPLANTATION INCLUDING HOW TO PROPERLY INSERT THE INFLOW CONDUIT. THE OPENING OF THE SEALED INFLOW CONDUIT SHOULD BE DIRECTED TOWARD THE MITRAL VALVE AND AWAY FROM THE INTERVENTRICULAR SEPTUM. ADDITIONALLY, PRIOR TO ADVANCING THE INLET EXTENSION INTO THE LEFT VENTRICLE THROUGH THE APICAL SEWING RING, THE USER IS INSTRUCTED TO REMOVE THE GLOVE TIP FROM THE SEALED INFLOW CONDUIT AND THE CENTERING FIXTURE FROM THE APICAL SEWING RING AND TO INSPECT THE VENTRICLE AND REMOVE ANY PREVIOUSLY FORMED CLOTS THAT MAY CAUSE EMBOLISM OR ANY TRABECULAE THAT MAY IMPEDE FLOW. THE IFU STATES THAT THE LOW FLOW HAZARD ALARM WILL BE TRIGGERED WHEN PUMP FLOW IS LESS THAN 2.5 LPM AND EXPLAINS THAT CHANGES IN PATIENT CONDITIONS CAN RESULT IN LOW FLOW. THIS IFU ALSO OUTLINES THE INDICATIONS OF DRIVELINE DAMAGE, AS WELL AS HOW TO RESPOND TO SUCH EVENTS. THE SECTION ENTITLED ¿ALARMS AND TROUBLESHOOTING¿ CONTAINS INFORMATION REGARDING ALARMS ON THE SYSTEM CONTROLLER AND THE PROPER ACTIONS ASSOCIATED WITH THEM. NO FURTHER INFORMATION WAS PROVIDED. THE MANUFACTURER IS CLOSING THE FILE ON THIS EVENT.
THE PATIENT WAS IMPLANTED WITH A LEFT VENTRICULAR ASSIST DEVICE (LVAD) ON (B)(6) 2011. IT WAS REPORTED THE PATIENT UNDERWENT A ROUTINE TRANSPLANT. DURING EVALUATION OF THE DEVICE, A THROMBUS WAS FOUND WITHIN THE DEVICE AS WELL AS DAMAGE TO THE BIONATE AND SILICONE JACKET LAYERS OF THE DRIVELINE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 329932 | HEARTMATE II LEFT VENTRICULAR ASSIST SYSTEM | LEFT VENTRICULAR ASSIST SYSTEM | DSQ | THORATEC CORPORATION | 104911 | 00813024011170 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 53 YR | Other |